background image

107801 - Rev B

RL-DBS

Digital Baby Scale

Operation Manual

Disponible

en Español 

Visite 

ricelake.com/spanish

para ver todos los ma

teriales

RLWS disponibles en Español

Содержание RL-DBS

Страница 1: ...107801 Rev B RL DBS Digital Baby Scale Operation Manual Disponible en Español Visite ricelake com spanish para ver todos los materiales RLWS disponibles en Español ...

Страница 2: ......

Страница 3: ... through Rice Lake Weighing Systems Course descriptions and dates can be viewed at www ricelake com or obtained by calling 715 234 9171 and asking for the training department Technical training seminars are available through Rice Lake Weighing Systems Course descriptions and dates can be viewed at www ricelake com or obtained by calling 715 234 9171 and asking for the training department 2015 Rice...

Страница 4: ...ii RL DBS Operation Manual Rice Lake continually offers web based video training on a growing selection of product related topics at no cost Visit www ricelake com webinars ...

Страница 5: ...s included 120V or 230V AC adapter included Automatic shut off ensures long battery life 1 1 Safety With the RL DBS 2 Digital Baby scale there are certain precautions that should be taken to prevent injury to the baby and damage to your scale 1 1 1 Safety Signals Safety Signal Definitions Indicates a potentially hazardous situation that if not avoided could result in death or serious injury Indica...

Страница 6: ...the infant is on the weighing tray Do not drop the scale or subject it to violent shocks For accurate weighing the scale must be placed on a flat stable surface For accurate weighing verify proper operation according to the procedure described in this manual before each use Operation at other voltages and frequencies than specified could damage your equipment Utilizing adapters not supplied by Ric...

Страница 7: ...pter The AC power adapter plugs into the back of the indicator as shown below in Figure 2 1 Figure 2 1 AC Power Connection Site The RL DBS Digital Baby scale is also capable of running on four 4 AA batteries if no additional power source is available If the Lo prompt is showing on the display replace the battery or connect the scale to an AC power source as soon as possible for accurate weighing F...

Страница 8: ...all key and see the baby s weight again Negative or Minus indicator When the scale is on this display shows a negative weight reading zero ZERO indicator When the scale is on this display indicates that the scale is stable and ready for use mem Memory indicator o Motion indicator This is displayed when a weight reading is complete It is not displayed while the weight reading is being calculated 6 ...

Страница 9: ... the RECALL key If the setup procedure failed refer to the troubleshooting section of this manual If the problem is not resolved refer to a qualified service provider 4 0 Scale Operation with Tare The RL DBS Digital Baby scale can accurately weigh babies up to 44 Lb 20 Kg Use the following steps to weigh a baby 1 Press the ON TARE key to turn the scale on 2 Wait until the 0 0 0 if set up in Lb Oz ...

Страница 10: ...h the scale Remove the infant interfering object from the weighing tray from the scale Display did not show 0 0 before weighing Remove the baby zero the scale and begin the weighing process again Scale is not placed on a level surface Place the scale on a stable surface and begin the weighing process again The weighing tray is not placed properly Place the weighing tray in its proper place Scale i...

Страница 11: ...raying or for broken or bent prongs 6 2 Cleaning Proper care and cleaning is essential to ensure a long life of accurate and effective operation Before beginning the cleaning process disconnect the scale from the AC power source 1 Clean all external surfaces with a clean damp cloth or tissue Mild soap and water solution may be used Dry with a clean soft cloth 2 Do not immerse the scale into cleani...

Страница 12: ...d was not caused by accident misuse neglect alteration improper installation improper repair or improper testing RLWS shall be the sole judge of all alleged non conformities Such equipment has not been modified altered or changed by any person other than RLWS or its duly authorized repair agents RLWS will have a reasonable time to repair or replace the defective equipment Buyer is responsible for ...

Страница 13: ......

Страница 14: ...68 USA U S 800 472 6703 Canada Mexico 800 321 6703 International 715 234 9171 Europe 31 0 26 472 1319 www ricelake com www ricelake mx www ricelake eu www ricelake co in m ricelake com Rice Lake Weighing Systems 03 2015 PN 107801 Rev B ...

Отзывы: