Minimum weight setting for measurement
Display settings: HOME key
APP button
[Application] settings screen
Application selection button Select [Minimum weight alert function]
HOME key
Weighing screen with the minimum weight alert function
Minimum weight setting button
Input by measuring repeatability button [Minimum weight setting] screen
1
Measurement method
2
Measurement tolerance
3
Measure button
Name
Setting value
Description
1
Minimum weight
measurement
method
Automatic
(internal weight)
Manual (external
weight)
Select the minimum weight measurement method.
2
Minimum weight
measurement
tolerance
0.1 %
.
1.0 %
.
Select the minimum weight measurement tolerance.
3
Measure button
-
Start the minimum weight measurement.
When the measurement is complete, the [Minimum weight
measurement result] screen described in “
” is displayed.
Settings in the red box are default values (factory settings).
This is the screen for measuring the minimum weight.
Automatic (internal weight)
For this setting, the minimum weight is automatically measured with the internal weight in the
[Repeatability measurement mode] screen. When the measurement is complete, the [Minimum weight
measurement result] screen described in “
5.5.4. Minimum weight measurement result
” is displayed.
1
Back button
Name
Description
1
Back button
Returns to the previous screen.
Содержание BA-T Series
Страница 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Страница 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Страница 107: ...107 Output example for daily check output results Daily check output results Daily check PDF results...
Страница 117: ...117 Output example for periodic check output results Periodic check output results...
Страница 118: ...118 Periodic check results PDF...
Страница 129: ...129 AND MEET analysis graph example...
Страница 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...
Страница 219: ......