95
10. [Sensitivity adjustment] screen
Display the [Sensitivity adjustment] screen as follows:
Display settings: MENU key
Sensitivity adjustment button
[Sensitivity adjustment] screen
1
Int. sensitivity adjustment button
3
Automatic sensitivity adjustment button
2
Ext. sensitivity adjustment button
4
Sensitivity adjustment setting button
Name
Description
1
Int. sensitivity adjustment button
Displays the [Internal sensitivity adjustment] screen to execute the internal
sensitivity adjustment.
After adjustment, the [Sensitivity adjustment result] screen for the internal
sensitivity adjustment is displayed.
2
Ext. sensitivity adjustment button
Displays the [External sensitivity adjustment] screen to start the external
sensitivity adjustment.
After adjustment, the [Sensitivity adjustment result] screen for the external
sensitivity adjustment is displayed.
3
Automatic sensitivity adjustment
button
Displays the [Automatic sensitivity adjustment] settings screen.
The automatic sensitivity adjustment is executed.
4
Sensitivity adjustment setting
button
Displays the [Sensitivity adjustment setting] screen. You can configure the
settings for the CAL button, standard value for external weight values and
internal weight value correction for the weighing screen.
Since the balance's resolution is high, weighing values may change due to gravity and daily
environmental changes. It is necessary to perform sensitivity adjustment with the weight in order to keep
the weighing values from changing even if gravity or the environment changes.
It is recommended that you perform sensitivity adjustment if the balance is installed for the first time or
relocated, or when the weighing values change significantly in daily check, etc.
Sensitivity adjustment means to adjust the weighing value of the balance using the reference weight or
internal weight.
Caution on sensitivity adjustment
Do not allow vibration, drafts or temperature changes to affect the balance especially during sensitivity
adjustment.
The GLP/GMP (etc.) compliant report can be output in sensitivity adjustment. A computer or optional
printer is required for GLP output. A timestamp (clock and calendar) is available for the GLP output
using the clock function of the balance.
You can set the clock function from the [Date/time setting] screen.
Display settings: MENU key
System settings button
Date/time button
[Date/time setting] screen
Содержание BA-T Series
Страница 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Страница 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Страница 107: ...107 Output example for daily check output results Daily check output results Daily check PDF results...
Страница 117: ...117 Output example for periodic check output results Periodic check output results...
Страница 118: ...118 Periodic check results PDF...
Страница 129: ...129 AND MEET analysis graph example...
Страница 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...