205
Error Display Error Code
Description and possible countermeasure
Weighing display E
Overload error
A sample beyond the balance weighing capacity has been placed on the pan.
Remove the object from the pan.
Weighing display -E
Weighing pan error
The weighing value is too light. Check that the weighing pan is installed
correctly.
Set the weighing pan correctly. Perform sensitivity adjustment.
Weighing display Lo
Sample mass error
The sample is too light to be stored as a sample mass for the counting mode or
percent mode.
rtc PF
Clock battery error
The clock backup battery has been depleted. Set the time and date. Even if the
clock backup battery is depleted, the clock and calendar function works normally
as long as the balance is powered with the AC adapter. If this error appears
frequently, contact your local dealer.
LoWVoLt
Power supply voltage fault
The voltage supplied from the AC adapter is abnormal.
Check that the AC adapter is the one supplied with the balance (refer to “
SD Error
Repeatability error
SD Error
The standard deviation (SD) of repeatability exceeded 50 d. Review the
installation environment of the balance.
"d" is a unit of readability.
Содержание BA-T Series
Страница 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Страница 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Страница 107: ...107 Output example for daily check output results Daily check output results Daily check PDF results...
Страница 117: ...117 Output example for periodic check output results Periodic check output results...
Страница 118: ...118 Periodic check results PDF...
Страница 129: ...129 AND MEET analysis graph example...
Страница 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...
Страница 219: ......