80
Impact shock detection (ISD) function
This function detects impact shocks to the mass sensor section, displays the impact level, and records
it. (Available with display unit software versions 1.010 or later)
By lowering the impact level at the time of loading, it is possible not only to alleviate variation in the
weighing value but also to reduce the risk of failure of the mass sensor section. Especially when
incorporating the balance in a production line, etc. and weighing by means such as an automated
system, impact greater than expected may be applied to the sensor. When designing automatic
systems and the like, it is recommended that you minimize the impact level as much as possible while
checking the shock indicator.
Impacts of impact level 3 or higher are stored on the balance with date and time.
For details, refer to “
13.29 Impact shock detection history
Impact shock detection history”.
You can hide the impact level by turning off the Warning display button.
If the impact level is level 3 or higher, the record will be kept automatically even if the impact level is
hidden.
Caution
Impact on the weighing sensor is not only that applied to the weighing pan when loaded, but also may
be impact applied from the table on which the balance is installed. The impact detection function also
works for impact coming from the table.
The shock indicator has 5 levels from level 0 to level 4.
Impact level
Shock
indicator
Buzzer
Contents
Level 0
No indicator
No beeps
Safe
Level 1
Caution
Level 2
Caution:
Alleviate impact shocks
Level 3
One beep
Warning:
Do not apply any more impact shocks
Level 4
Two beeps
Danger:
Sensor may be damaged
Содержание BA-T Series
Страница 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Страница 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Страница 107: ...107 Output example for daily check output results Daily check output results Daily check PDF results...
Страница 117: ...117 Output example for periodic check output results Periodic check output results...
Страница 118: ...118 Periodic check results PDF...
Страница 129: ...129 AND MEET analysis graph example...
Страница 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...
Страница 219: ......