203
23. Error Codes
Error code list
Error Display Error Code
Description and possible countermeasure
EC, E00
Communications error
A protocol error occurred in communications.
Check the format, baud rate, and parity.
EC, E01
Undefined command error
An undefined command was found.
Check the transmitted command.
EC, E02
Not ready
The received command cannot be executed:
(e.g.) Q command was received when not in weighing mode.
(e.g.) Q command was received while re-zeroing.
Adjust the delay time to transmit a command.
EC, E03
Timeout error
With the command timeout setting, there was a wait time of approximately 1
second or more while receiving command characters. Check the
communication.
EC
,
E04
Character length error
The number of characters in the received command has exceeded the limit.
Check the command to transmit.
EC
,
E06
Format error
The description of the received command is incorrect:
(e.g.) The number of digits of numerical values is incorrect.
(e.g.) There are alphabet characters among the numerical values.
Check the transmitted command.
EC
,
E07
Parameter setting error
The value of the received command has exceeded the allowed value.
Check the setting range of the numerical value of the command.
Error 1
EC, E11
Stability error
Weighing value is unstable and therefore the “zero display”, “sensitivity
adjustment”, “calibration test”, etc. cannot be executed. Check around the pan.
Improve the environment of the installation location to prevent factors such as
vibration, draft, temperature change, static electricity and magnetic field from
influencing the balance.
Wait for 10 seconds to clear the error.
Error 2
Out of the setting range
The value to be set exceeds the setting range.
Set again within the setting range.
Error 3
Malfunction of the internal memory element of the balance
If this error continues to be displayed, repair is necessary. Please contact your
local dealer for repair.
Содержание BA-T Series
Страница 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Страница 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Страница 107: ...107 Output example for daily check output results Daily check output results Daily check PDF results...
Страница 117: ...117 Output example for periodic check output results Periodic check output results...
Страница 118: ...118 Periodic check results PDF...
Страница 129: ...129 AND MEET analysis graph example...
Страница 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...
Страница 219: ......