background image

Instruction Manual

USS-DBS81 Analytical Balance

Содержание USS-DBS81

Страница 1: ...Instruction Manual USS DBS81 Analytical Balance ...

Страница 2: ...n and Function 3 2 1 Components 3 2 2 Keyboard 4 2 3 Display Panel 5 Unpacking 6 Before Using 7 4 1 Leveling the balance 7 4 2 Calibration 8 Operation 9 5 1 Basic Weighing 9 5 2 Unit Switching 10 5 3 Counting 10 Setting 12 6 1 Minimum weighing 12 Technical Data 13 ...

Страница 3: ...an used but also for special use scenarios that single screen ordinary balances can not meet U S Solid s dedicated customer service staff are available to answer any inquiries regarding applications and accessories Please read the manual completely and follow the usage instructions before installation and operation as this will help you to make full use of the functions and performance of the USS ...

Страница 4: ...er Use a power source voltage frequency outlet type adapted to the specification of the balance If excessive voltage is used the balance may overheat and be damaged or cause a fire Operate the balance on a stable rigid and flat table Handle the balance carefully It is a precision device subjecting it to impact may result in a malfunction If the device is not be used for a long period of time the p...

Страница 5: ... Components 1 Upper Handle 2 Upper Glass 3 Housing Ⅰ 4 Vertical Strut 5 Rear Glass 6 Side Glass 7 Side Handle 8 Weighing Pan 9 Glass 10 Tray 11 Housing Ⅱ 12 Housing Ⅲ 13 Housing Ⅳ 14 Leveling foot 15 Dust Cover 16 Battery Cover ...

Страница 6: ...rn the power on or off Enter the counting function Long press with TARE to enter the minimum weight setting Tare the balance Long press with SET to enter the minimum weight setting Performs weight calibration Switch the weighing units ...

Страница 7: ...tical Balance provides dual display screens one front and one rear No Description 1 Indicates positive values Indicates negative values Stabilization indicator Disappearance indicates stabilization of balance 2 Weight value 3 Weight unit ...

Страница 8: ...g accessories are part of the standard equipment for your new USS DBS81 Analytical Balance 1 Balance Main Body 1 Weighing Pan 1 Draft Shield 1 Power Cable 1 Calibration Weight 1 Instruction Manual 1 Quality Certification Check the instrument for damage in transit Immediately inform the U S Solid customer service if you have any complaints or parts are missing ...

Страница 9: ...on please follow the calibration method below to calibrate the balance 4 1 Leveling the balance Accurate horizontal positioning necessary for repeatable measurements and exact results To compensate for small irregularities or tilts at this location the instrument needs to be leveled Adjust the leveling feet of the analytical balance until the air bubble in the indicator is centered The level indic...

Страница 10: ...Changes in ambient conditions Prior to each use 1 Zero the balance in the weighing state 2 Press the CAL button and CAL will be displayed 3 After that 200 000 or 100 000 will be flashing displayed then place the calibration weight on the pan 4 will be displayed then 200 000 or 100 000 will be displayed again 5 Remove the calibration weight and will again displayed 6 The calibration is completed an...

Страница 11: ...Weighing a sample that needs to be in a container 1 Open the glass door of the weighing chamber and place the weighing container on the weighing pan then close the glass door again 2 Wait for the display to stabilize Then press O T to tare and the displayed value will return to zero 3 Open the glass door and place the sample s to be weighed in the weighing container then close the glass door 4 Aft...

Страница 12: ...nting function that meets various industrial counting requirements especially the counting function for small components To ensure the accuracy of the counting function for small components it is necessary to ensure that the weight of all samples is consistent and that the weight of a single object is 0 004g 1 Press SET and COU PCS will be displayed 2 After that 10 PCS will be flashing displayed 3...

Страница 13: ...e 10 PCS again displayed and the corresponding number appears Exit counting function 1 Press SET and will be displayed 2 After that ERR will be flashing displayed 3 Press SET to exit the counting function and return to the weighing interface ...

Страница 14: ...m weight can be set according to the use requirements 1 Long press TARE and SET simultaneously until 1 d displayed 2 Press SET or CAL to turn up or turn down respectively from 1d to 5d d means minimum scale value 0 001 g 3 Press TARE to confirm ...

Страница 15: ...bration Weight 100 g 200 g 200 g Readability 0 001g 1 mg Repeatability 0 002 g Linearity Deviation 0 003 g Stabilization Time Within 3 seconds Pan Size Φ3 55 inches 90 mm Interface RS232 Display Screen Dual Power 110V 60Hz AC Dimensions 10 6 7 9 10 6 inches Net Weight 7 72 lbs Shipping Weight 8 82 lbs ...

Страница 16: ...Contact Feel free to visit our website www ussolid com You can email us at service ussolid com You can call one of our friendly customer service representative at 1 800 209 4177 ...

Отзывы: