32
Zero-point, tare, and weighing range
Entering the weighing mode
The balance will determine the reference zero point when the ON:OFF key
is pressed to enter the
weighing mode.
Depending on the load condition at that time, the balance automatically judges whether to set the zero-
point or to tare.
The condition for determining which is used is the "power-on zero range", and when the power-on zero
range is exceeded, the tare operation is performed.
Re-zero operation
By pressing the RE-ZERO button
, the displayed value can be set to zero.
The re-zero operation with the RE-ZERO button
will automatically judge whether to set the zero-
point or to tare.
Weighing range
The weight range that the balance can display varies depending on the model. When the gross weight
weighed exceeds the maximum display, E is displayed to indicate that the weighing range is
exceeded. When exceeded in the negative direction, –E is displayed.
Gross weight = Net weight (weighing value minus tare) + Tare weight
Model
Power-on zero range
Zero range
-E display range
BA-6TE
BA-6DTE
Approx. ±0.6 g
Approx. -0.6 g to +0.1 g
Approx. -0.6 g
BA-225TE
BA-225DTE
Approx. ±22 g
Approx. -22 g to +4.4 g
Approx. -22 g
BA-125DTE
Approx. ±12 g
Approx. -12 g to +2.4 g
Approx. -12 g
Содержание BA-T Series
Страница 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Страница 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Страница 107: ...107 Output example for daily check output results Daily check output results Daily check PDF results...
Страница 117: ...117 Output example for periodic check output results Periodic check output results...
Страница 118: ...118 Periodic check results PDF...
Страница 129: ...129 AND MEET analysis graph example...
Страница 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...
Страница 219: ......