186
If the gross/tare weight is added in the [Data to be added] screen of the function table, the second
header corresponding to the output data is appended to the header.
S T
,
N
,
+ 0 1
.
2 3 4 5 6 7
g
C
R
L
F
2nd header
Data
Unit
Terminator
N
Net weight
C
R
:
Carriage return,
ASCII 0Dh
G
Gross weight
L
F
:
Line feed,
ASCII 0Ah
T
Tare weight
:
Space,
ASCII 20h
P T
Preset tare weight
Display settings: MENU key
System settings button
Communication button
Data output button
Data to be added button
[Data to be added] screen
DP format (dump print)
This format is suitable for dump printing.
Consists of 16 characters (not including the terminator).
The condition of the data is indicated with a 2-character header.
The polarity sign is added right before the value if it is not an overload or zero.
The data is zero-suppressed (leading zeros are replaced with spaces).
The unit consists of three characters.
W T
+ 1
.
2 3 4 5 6 7
g
C
R
L
F
Header
Data
Unit
Terminator
W T
When stable
C
R
:
Carriage return,
ASCII 0Dh
U S
When unstable
L
F
:
Line feed,
ASCII 0Ah
Q T
Counting mode when
stable
:
Space,
ASCII 20h
KF format
This is the Karl-Fischer moisture meter format.
Consists of 14 characters (not including the terminator).
There are no headers.
The polarity sign is added to the first character if it is not an overload or zero.
The data is zero-suppressed (leading zeros are replaced with spaces).
When stable, the unit is output. When not stable, the unit is not output.
+
1
.
2 3 4 5 6 7
g
C
R
L
F
Data
Unit
Terminator
C
R
:
Carriage return,
ASCII 0Dh
g
With a unit when stable
L
F
:
Line feed,
ASCII 0Ah
No unit when unstable
:
Space,
ASCII 20h
Содержание BA-T Series
Страница 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Страница 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Страница 107: ...107 Output example for daily check output results Daily check output results Daily check PDF results...
Страница 117: ...117 Output example for periodic check output results Periodic check output results...
Страница 118: ...118 Periodic check results PDF...
Страница 129: ...129 AND MEET analysis graph example...
Страница 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...