204
Error Display Error Code
Description and possible countermeasure
Error 6
EC
,
E16
Internal weight error
Applying the internal weight does not yield a change in the mass value as
specified.
Confirm that there is nothing on the pan and perform the operation from the
beginning.
Error 7
EC
,
E17
Internal weight error
The internal weight application mechanism does not function properly. Perform
the operation from the beginning.
Error 8
Abnormality in the internal memory data of the balance
If this error continues to be displayed, repair is necessary. Please contact your
local dealer for repair.
Error 9
Abnormality in the internal memory data of the balance
If this error continues to be displayed, repair is necessary. Please contact your
local dealer for repair.
Error 16
Internal weight error
Applying the internal weight did not yield a weight change.
Make sure that there is nothing on the weighing pan and perform the same
operation again.
Error 17
Internal weight error
The internal weight application mechanism does not function properly.
Perform the operation from the beginning.
Error 20
Sensitivity adjustment weight error (heavy)
The weight is too heavy. Check the nominal value of the weight.
Send the CAL command or wait for 5 seconds to clear the error.
Error 21
Sensitivity adjustment weight error (light)
The weight is too light. Check the nominal value of the weight.
Send the CAL command or wait for 5 seconds to clear the error.
CAL E
EC
,
E20
Sensitivity adjustment weight error (Positive value)
The weight is too heavy.
Check around the pan. Check the mass value of the weight.
-CAL E
EC
,
E21
Sensitivity adjustment weight error (Negative value)
The weight is too light.
Check around the pan. Check the mass value of the weight.
Con Err
Weighing unit error
Data from the weighing unit has not been received.
Check the connection between the weighing unit and the display unit, and turn
the power off and then on again.
Door Err
Breeze break error
Communication with the breeze break is not possible.
Unplug the power supply and reconnect the cable between the breeze break
unit and the weighing unit.
S/N Error
Serial number error
The serial numbers of the weighing unit and the display unit do not match.
Reconnect with the correct combination.
Содержание BA-T Series
Страница 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Страница 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Страница 107: ...107 Output example for daily check output results Daily check output results Daily check PDF results...
Страница 117: ...117 Output example for periodic check output results Periodic check output results...
Страница 118: ...118 Periodic check results PDF...
Страница 129: ...129 AND MEET analysis graph example...
Страница 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...
Страница 219: ......