9. Sensitivity Test
The sensitivity test is performed with the selected loading points.
Follow the instruction for measurement operations.
The setting in the [Sensitivity test] screen is common to the standard value setting for the daily/periodic
check settings.
When the test is complete, the [Sensitivity test result] screen is automatically displayed.
7
8
1
9
2
10
3
4
11
5
6
Name
Setting value
(setting range)
Description
1
Settings display screen
-
Displays the set tolerance.
2
Number of loading points
for measurement
1 point ,
2 points
You can select the number of loading points for
measurement.
3
Measuring load 1
-
Displays the nominal value and tolerance.
4
Measuring load 2
-
Displays the nominal value and tolerance.
5
Measuring load display
-
Displays the measuring load.
6
Back button
-
Cancels and returns to the previous screen.
7
Settings button
-
Displays the [Standard value setting] screen.
Set the measurement count and tolerance (standard
deviation).
8
Measure button
-
Executes the measurement.
9
Sensitivity test result
display
-
Displays the sensitivity test result.
10
Skip button
-
Skips this check item and proceeds to the next item.
11
Next button
-
Proceeds to the next item.
Settings in the red box are default values (factory settings).
Содержание BA-T Series
Страница 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Страница 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Страница 107: ...107 Output example for daily check output results Daily check output results Daily check PDF results...
Страница 117: ...117 Output example for periodic check output results Periodic check output results...
Страница 118: ...118 Periodic check results PDF...
Страница 129: ...129 AND MEET analysis graph example...
Страница 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...
Страница 219: ......