Printing information other than weighing value data
To print sensitivity adjustment/calibration test reports (GLP compliant output) or to perform output of the
statistical calculation result calculated by the balance, change the setting of the printer to the dump print
mode.
Settings to print information other than weighing value data with the AD-8127
PRN .MODE
Description
DUMP
Dump print mode
Switching the AD-8127 print mode (PRN MODE)
By pressing and holding the ENTSAVE key of the AD-8127, it is possible to switch between EXT.KEY
(External key print mode) and DUMP (Dump print mode) without using the function table of the printer.
This is convenient when temporarily switching the AD-8127 to dump print mode for GLP output, etc.
The data output compliant with "GLP/GMP" can be output to a personal computer or optional printer.
Set the GLP/GMP settings in the following screen.
MENU key
System settings button
Communication button
GLP output button
Set in the [GLP output] screen.