207
Optimizing the Ionizer
Static elimination method can be optimized with the switches on the rear. Refer to the following table for
the functions of switches. Change the settings of switch 1 and switch 2 as shown in the table when
eliminating static electricity out of the effective range.
The factory settings of switches are all ON (upper side). It means static elimination method is in ‘timer
mode,’ ‘discharging time’ is for 3 seconds, and ‘built-in IR sensor’ and ‘buzzer’ are available.
Switch No. / Item
Switching location
Description
Switch 1
Static elimination method
ON Upper side
Timer mode
*1
OFF Lower side
Manual mode
*2
Switch 2
Discharging time
*3
ON Upper side
3 seconds
OFF Lower side
10 seconds
Switch 3
Built-in IR sensor
ON Upper side
Available
OFF Lower side
Not available
Switch 4
Buzzer
*4
ON Upper side
Available
OFF Lower side
Not available
*1
When the ‘timer mode’ is selected with switch 1, static elimination is performed for the ‘discharging time’
set with switch 2.
*2
When the ‘manual mode’ is selected with switch 1 and static elimination starts, it continues until the built-
in IR sensor or the optional IR switch performs again. The ‘manual mode’ does not stop static
elimination automatically.
*3
Discharging time is settable only when the static elimination mode is in the ‘timer mode’ (switch 1 is on).
*4
Buzzer rings when the power is turned on or whenever the IR sensor or IR switch performs.
Maintaining the ionizer
Caution: Do not touch electrode units while the ionizer is in operation. Doing so may result in electric shock.
Dust and other substances adhere to the electrode needle area of the ionizer over time, degrading the
static elimination capacity. To maintain the performance, clean the electrode needle on electrode units
with a dry cotton swab, etc. on a regular basis.
If the static elimination capacity does not recover because the tip of the electrode needle on an
electrode unit is worn out, replace all of the four electrode units with new ones. The lifetime of electrode
units is approx. 10000 hours.
Replacement procedure
Step 1. Disconnect the balance connection cable and turn off the power.
Step 2. Rotate electrode units counterclockwise by 45° to pull them out.
Step 3. Insert new electrode units and rotate them clockwise by 45° to secure them.
Electrode unit
Electrode needle Clean it
Step 2
Step 3
1 2 3 4
ON
The factory settings of switches are all ON.
Switch knob
ON
OFF
Содержание BA-T Series
Страница 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Страница 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Страница 107: ...107 Output example for daily check output results Daily check output results Daily check PDF results...
Страница 117: ...117 Output example for periodic check output results Periodic check output results...
Страница 118: ...118 Periodic check results PDF...
Страница 129: ...129 AND MEET analysis graph example...
Страница 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...
Страница 219: ......