177
17. Printing Weighing Value Data with a Printer
Refer to the following setting examples for the printer settings and the balance's function table according
to the printer to be used and the method of printing weighing data.
For an AD-8127 multi-functional compact printer
Printing only weighing value data
Common settings for the balance when printing only weighing value data with the AD-8127
Item
Description
Data format
A&D standard format
Settings when printing only weighing value data with the AD-8127
Printing method
Balance function table
AD-8127 function table
Description
PRN .MODE Description
Prints weighing value data
when the PRINT button of
the balance is pressed
Key mode: Output only when the display is stable
EXT.KEY
External
key print
mode
Key mode: Output regardless of whether the
display is stable or unstable
*1
Key mode: Output when stable
Automatically prints
weighing value data
according to the change in
the weighing value
Auto print mode: Reference = zero point
Auto print mode: Reference = Last stable value
Prints weighing value data
at regular intervals
Interval output mode
*1
Prints weighing value data
when the PRINT button of
the printer is pressed
Stream mode
*1
MANUAL
Manual
print mode
Prints weighing value data
in chart format.
CHART
Chart
print mode
*1
Unstable data is also output.
To set the AD-8127 to a mode other than dump print mode and also print unstable data, change the
setting to "Print unstable data ("US PRN", "PRINT")" in the function table of the AD-8127.
For details on the A&D standard format, refer to “
”.
You can select the A&D standard format for each connection in the following screen.
MENU key
System settings button
Communication button
RS-232C interface button
In the [RS-232C interface] screen, select [A&D standard format].
For details about [Key mode], [Auto print mode], [Interval output mode] and [Stream mode],
refer to “
”. Set the data output mode in the following screen.
MENU key
System settings button
Communication button
Data output button
Data output mode button
In the [Data output mode] screen,
select the mode.
Содержание BA-T Series
Страница 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Страница 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Страница 107: ...107 Output example for daily check output results Daily check output results Daily check PDF results...
Страница 117: ...117 Output example for periodic check output results Periodic check output results...
Страница 118: ...118 Periodic check results PDF...
Страница 129: ...129 AND MEET analysis graph example...
Страница 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...
Страница 219: ......