Command settings
Display settings: MENU key
System settings button
Communication button
Data output button
Command settings button
[Command settings] screen
1
Terminator
2
AK (acknowledge), error code
3
Command time out
4
Back button
Name
Setting value
(setting range)
Description
1
Terminator
C
R
L
F
, C
R
Selects the terminator for the output data.
2
AK (acknowledge),
error code
OFF , ON
Select the response (received, processing or process
completed) to all commands sent from PC or PLC.
3
Command time out
No limit,
Limit to one second Sets the command timeout.
4
Back button
-
Returns to the previous screen.
Settings in the red box are default values (factory settings).
AK
:
Acknowledgment, ASCII 06h
C
R
:
Carriage return, ASCII 0Dh
L
F
:
Line feed, ASCII 0Ah
When [AK (acknowledge), error code] is set to ON in the command settings, the balance always
responds to reception of all commands sent from a personal computer or a PLC. Checking the code that
is responded improves the reliability of the communication.
Balance response
When [AK (acknowledge), error code] is set to ON in the function table, the balance responds as
follows.
When a command requesting data is sent to the balance:
If the balance cannot send the requested data, it sends an error code (EC,Exx).
If the balance can output the requested data, it sends the requested data.
When a command controlling the balance is sent to the balance:
If the balance cannot execute the command, it sends an error code (EC, Exx).
If the balance can execute the command, it sends the AK code (acknowledgment, ASCII 06h).
Command
Description
ON
Turns the display on.
P
Turns the display on/off. (Only when the display is on.)
R, RZ
Same as the RE-ZERO button
.
T, TR
Same as the TARE button
.
ZR
Zero
If the load is within ±2% of the capacity from the initial zero point, the zero
point is updated, the tare value is cleared and the display is set to zero. If the
load exceeds ± 2%, no processing is done.
CAL
Sensitivity adjustment with the internal weight.
EXC
Sensitivity adjustment with an external weight.
Содержание BA-T Series
Страница 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Страница 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Страница 107: ...107 Output example for daily check output results Daily check output results Daily check PDF results...
Страница 117: ...117 Output example for periodic check output results Periodic check output results...
Страница 118: ...118 Periodic check results PDF...
Страница 129: ...129 AND MEET analysis graph example...
Страница 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...
Страница 219: ......