HPLC function
HPLC function usage
The HPLC function allows you to register in the recipe the sample to be weighed and the molarity
(mol/L, mmol/L) or the target value in unit of mass (g) and the tolerance for the target value, and to
perform weighing according to the recipe.
The target value (g) of the sample is calculated by the following formula.
Target value (g) = Molarity (mol/L) x Solution volume (L) x Molecular weight
Caution
Initializing the balance clears the sample contents and registered recipes (excluding those registered at
the time of shipment from the factory).
If you configure this display, the HOME screen changes to If you configure this display, the HOME
screen changes to HPLC mode.
Display settings: HOME key
APP button
[Application] settings screen
Application selection button Select [HPLC mode]
HOME key
[HPLC mode] screen
1
5
2
3
4
Name
Description
1
HPLC mode
Displayed when in the HPLC mode.
2
Got to the recipe list
button
Displays the [Recipe list].
Also displays the name of the currently selected recipe.
3
Skip button
Skips tare weighing and displays the [HPLC mode - Sample weighing] screen.
If the recipe is incomplete, the button is disabled.
4
SAVE button
Weighs the tare value and displays the [HPLC mode - Sample weighing] screen.
If the recipe is incomplete, the SAVE button
is disabled.
If the weighing value is negative, it is recorded as 0 g.
5
Display indicating that
the recipe is incomplete
If the selected recipe is incomplete:
The message “The recipe is incomplete. Please check the recipe details.” is
displayed and the Skip button
and SAVE button
are disabled.
Check the recipe name and the target value and tolerance of the sample
registered in the recipe.
If the selected recipe is complete:
The message “Please place item to be tared” is displayed and the Skip button
and SAVE button
are enabled.
This is the screen for [HPLC mode]. This mode can be used only with the unit “g”.
Re-zero will be applied automatically when the screen changes to [HPLC mode].
If the tare weight plus sample target value exceeds the capacity, the message “Overload error” is displayed
for the item indicated by
5
in the figure above and the SAVE button SAVE button
is disabled.
Содержание BA-T Series
Страница 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Страница 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Страница 107: ...107 Output example for daily check output results Daily check output results Daily check PDF results...
Страница 117: ...117 Output example for periodic check output results Periodic check output results...
Страница 118: ...118 Periodic check results PDF...
Страница 129: ...129 AND MEET analysis graph example...
Страница 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...
Страница 219: ......