Normal measuring mode LMG450
78
to see the real phase shift between the phases.
The voltage measured with the third channel is also inverted.
5.1.2
Group A/B tab
The settings in group A and B are identical, so they are just described once. Each setting
influences all channels which belong to the group!
Sync
Selects the signal which is used for synchronisation of all channels belonging to this
group. There are following possible settings:
U
The voltage signal is used (channel 1 in group A and channel 4 in B)
I
The current signal is used (channel 1 in group A and channel 4 in B)
X
Extended Trigger. See Xtrig.
Line
The line signal is used
Extn
The signal at the external synchronisation jack is used.
Depending on the value of Sync there is one softkey which changes from Xtrig
(setting ‘X’) to Coupl (all other settings).
Coupl
Selects how the voltage or current signal is coupled to the following trigger stage.
This setting has no influence on the measured signals!
AC+DC
The signal is directly coupled, including all signal parts.
LP<80Hz
Low Pass Filter. Only the signal components below 80Hz are used
LP<300Hz
Low Pass Filter. Only the signal components below 300Hz are used. This is
very useful, when measuring frequency converters with fundamental
frequencies below 300Hz and clock frequencies above 300Hz.
HP>10Hz
High Pass Filter. Only the signal components above 10Hz are used. This is
useful to reject a DC component in the signal.
HP>30Hz
High Pass Filter. Only the signal components above 30Hz are used. This is
useful to reject a DC component in the signal.
BP 10-300Hz
Band Pass Filter. Only the signal components between 10Hz and 300Hz are
used.
BP 30-80Hz
Band Pass Filter. Only the signal components between 30Hz and 80Hz are
used. This is useful when synchronising to line frequent signals with very
much distortion.
AM 10-300Hz
Amplitude (De)Modulation. The signal will be demodulated when measuring
AM signals. Only the envelope between 10Hz and 300Hz is used. This is
useful, when measuring electronic gears or transformers.
Xtrig
Here you reach a menu where you can define very precise, what should be your
trigger condition. This menu should only be used from very experienced users,
because if you select wrong conditions, you might get wrong measuring results.
Summary of Contents for LMG450
Page 1: ...4 Channel LMG450 Power Meter User manual Status 2009 02 17...
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 8: ......
Page 10: ......
Page 18: ......
Page 22: ......
Page 36: ......
Page 76: ......
Page 102: ......
Page 112: ......
Page 118: ......
Page 124: ......
Page 230: ......
Page 240: ......
Page 292: ......