Z E S ZIMMER Electronic Systems GmbH Chapter 10
151
10.2.4.1.5.3 :PHWave?
FLPH?
Phw
SCPI:
:FETCh[:SCALar]:FLICker:PHWave? /qonly/ <list> |
:READ[:SCALar]:FLICker:PHWave? /qonly/ <list>
SHORT: FLPH? /qonly/ <list>
ID:
Phw
Mode:
Flicker
Type:
float
Suffix:
1...4
Value:
n/a
List:
0...31
Unit:
W
Group:
n/a
Reads the half wave power values. After each measuring cycle over 16 periods you can get 32
values. To get a correct value it is necessary to measure the current with the current channel
and not the voltage of a source!
10.2.4.1.5.4 :SOURce
Selects the source’s measuring results. They are measured at the current input jacks.
:CALCulate
:DISPlay
:FETCh
→
[:SCALar]
→
:CURRent
:FORMat
:CYCLe
:INITiate
:DINPut
:INPut
:ENERgy
:INSTrument
:FLICker
→
[:EUTest]
:MEMory
:FREQuency
:LTRemain
:READ
→
:HARMonics
:PHWave
:SENSe
:POWer
:SOURce
→
:APMoment
:SOURce
:RESistance
:STATe
:DC
:STATus
:SSYStem
:STRemain
:DELTat
:SYSTem
:VARiable
:DMAX
:TRIGger
[:VOLTage]
:HWTRms
:PLT
:PMOMentary
:PST
:RESult
10.2.4.1.5.4.1 :APMoment?
FSMO?
Pmoms
SCPI:
:FETCh[:SCALar]:FLICker:SOURce:APMoment? /qonly/ |
:READ[:SCALar]:FLICker:SOURce:APMoment? /qonly/
SHORT: FSMO? /qonly/
ID:
Pmoms
Mode:
Flicker
Type:
float
Suffix:
1...4
Value:
n/a
List:
n/a
Unit:
n/a
Group:
n/a
Reads the averaged momentary flicker level of the source. It is averaged over 16 periods.
10.2.4.1.5.4.2 :DC?
FSDC?
dcs
SCPI:
:FETCh[:SCALar]:FLICker:SOURce:DC? /qonly/ |
:READ[:SCALar]:FLICker:SOURce:DC? /qonly/
SHORT: FSDC? /qonly/
Summary of Contents for LMG450
Page 1: ...4 Channel LMG450 Power Meter User manual Status 2009 02 17...
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 8: ......
Page 10: ......
Page 18: ......
Page 22: ......
Page 36: ......
Page 76: ......
Page 102: ......
Page 112: ......
Page 118: ......
Page 124: ......
Page 230: ......
Page 240: ......
Page 292: ......