Z E S ZIMMER Electronic Systems GmbH Chapter 10
135
10.2.1.14 *WAI
*WAI
SCPI:
*WAI/nquery/
SHORT: *WAI/nquery/
ID:
n/a
Mode:
all
Type:
n/a
Suffix:
n/a
Value:
n/a
List:
n/a
Unit:
n/a
Group:
n/a
Waits until all pending selected device operations have been finished. Note: The instrument
handles commands in a queue, so when executing the *WAI all previous commands have been
executed. Thus the instrument is doing nothing when receiving the *WAI command. It has
been implemented to follow the standard IEEE488.2.
10.2.2 :CALCulate commands
Here you find commands which influence the script editor (formulas) or limits.
:CALCulate
→
:ENVironment
:DISPlay
:FORMula
:FETCh
:LIMit
:FORMat
:INITiate
:INPut
:INSTrument
:MEMory
:READ
:SENSe
:SOURce
:STATus
:SYSTem
:TRIGger
10.2.2.1.1 ENVironment
ENV
Env
SCPI:
:CALCulate:ENVironment <NRf>[,<NRi>]
SHORT: ENV <NRf>[,<NRi>]
ID:
Env
Mode:
all
Type:
float
Suffix:
n/a
Value:
n/a
List:
n/a
Unit:
n/a
Group:
optional [,<NRi>]: 0=Env0, ... ,7=Env7
Sets or queries an environment variable.
10.2.2.2 :FORMula
:CALCulate
→
:ENVironment
:DISPlay
:FORMula
→
[:DEFine]
:FETCh
:LIMit
:FORMat
:INITiate
:INPut
:INSTrument
:MEMory
Summary of Contents for LMG450
Page 1: ...4 Channel LMG450 Power Meter User manual Status 2009 02 17...
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 8: ......
Page 10: ......
Page 18: ......
Page 22: ......
Page 36: ......
Page 76: ......
Page 102: ......
Page 112: ......
Page 118: ......
Page 124: ......
Page 230: ......
Page 240: ......
Page 292: ......