No.
Name
Possible reason; what to do
-124
Too many digits
Number has more than 9 digits
-131
Invalid suffix
Suffix too big or small
-150
String data error
A’“’ is missing
-221
Settings conflict
Setting at the moment impossible. For example to
change a measuring range while autorange is active
-222
Data out of range
Happens for example at invalid <list> entries
-224
Illegal parameter value
Happens for example if you want to change to the (not
existing) measuring mode 27
other
Illegal error, please inform
your supplier
An error in the internal error handling. Please contact
ZES
:CALCulate
:DISPlay
:FETCh
:FORMat
:INITiate
:INPut
:INSTrument
:MEMory
:READ
:SENSe
:SOURce
:STATus
:SYSTem
→
:BEEPer
:TRIGger
:DATE
:ERRor
→
:ALL
:HELP
:COUNt
:KEY
[:NEXT]
:LANGuage
:OPTions
:PHEADER
:TIME
:VERSion
10.2.13.3.1 :ALL?
ERRALL?
SCPI:
:SYSTem:ERRor:ALL? /qonly/
SHORT: ERRALL? /qonly/
ID:
n/a
Mode:
All
Type:
n/a
Suffix:
n/a
Value:
n/a
List:
n/a
Unit:
n/a
Group:
n/a
Reads all errors, including error code and error description, separated by colons out of the
error/event queue.
10.2.13.3.2 :COUNt?
ERRCNT?
SCPI:
:SYSTem:ERRor:COUNt? /qonly/
Summary of Contents for LMG450
Page 1: ...4 Channel LMG450 Power Meter User manual Status 2009 02 17...
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 8: ......
Page 10: ......
Page 18: ......
Page 22: ......
Page 36: ......
Page 76: ......
Page 102: ......
Page 112: ......
Page 118: ......
Page 124: ......
Page 230: ......
Page 240: ......
Page 292: ......