Interfaces (IEEE option L45-O1) LMG450
154
10.2.4.1.5.4.10 :RESult?
FSRE?
SrcOK
SCPI:
:FETCh[:SCALar]:FLICker:SOURce:RESult? /qonly/ |
:READ[:SCALar]:FLICker:SOURce:RESult? /qonly/
SHORT: FSRE? /qonly/
ID:
n/a
Mode:
Flicker
Type:
long int
Suffix:
1...4
Value:
n/a
List:
n/a
Unit:
n/a
Group:
n/a
Reads the result of the flicker measuring at the source.
Bit 0: Set if the total evaluation of the flicker fails (=if any of the sub evaluation fails). Cleared
otherwise.
Bit 1: Set if at least one P
st
value was >1.0, cleared otherwise.
Bit 2: Set if the P
lt
value was >0.65 at the END of the measuring interval, cleared otherwise.
Bit 3: Set if d
max
was bigger than limit, cleared otherwise.
Bit 4: Set if d(t) was >3% for more than allowed time, cleared otherwise.
Bit 5: Set if d
c
was > 3%, cleared otherwise.
10.2.4.1.5.5 :STATe?
FSTA?
SCPI:
:FETCh[:SCALar]:FLICker:STATe? /qonly/ | :READ[:SCALar]:FLICker:STATe? /qonly/
SHORT: FSTA? /qonly/
ID:
n/a
Mode:
Flicker
Type:
long int
Suffix:
n/a
Value:
n/a
List:
n/a
Unit:
n/a
Group:
n/a
Reads the state of the flicker measuring.
0:
Reset
1:
Wait
2:
Run
3:
Stop
10.2.4.1.5.6 :STRemain?
FSTR?
SCPI:
:FETCh[:SCALar]:FLICker:STRemain? /qonly/ |
:READ[:SCALar]:FLICker:STRemain? /qonly/
SHORT: FSTR? /qonly/
ID:
n/a
Mode:
Flicker
Type:
long int
Suffix:
n/a
Value:
n/a
List:
n/a
Unit:
s
Group:
n/a
Reads the remaining short time for the actual short term measurement.
Summary of Contents for LMG450
Page 1: ...4 Channel LMG450 Power Meter User manual Status 2009 02 17...
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 8: ......
Page 10: ......
Page 18: ......
Page 22: ......
Page 36: ......
Page 76: ......
Page 102: ......
Page 112: ......
Page 118: ......
Page 124: ......
Page 230: ......
Page 240: ......
Page 292: ......