Z E S ZIMMER Electronic Systems GmbH Chapter 13
257
20kHz
0mV
100kHz
0mV
Voltage inputs for current measuring with isolating current sensors
Rated range value / V
0.12
0.25
0.5
1
2
4
Measurable TRMS value / V
0.15
0.3
0.6
1.2
2.5
5
Permissible peak value / V
0.25
0.5
1
2
4
8
Overload capability
100V continuously, 250V for 1s
Input resistance
100k
Ω
, 10pF
Common mode error no error per definition, because this input is not isolated from chassis
ground, so you can’t apply a common mode signal.
13.3.3 Accuracy
An example, how to handle this values, you find in 12.1.1, ‘Accuracy of measured and
computed values’
Measuring accuracy
The values are in
±
(% of measuring value + % of measuring range)
Frequency/Hz
DC
1Hz-1kHz 1kHz-5kHz 5kHz-20kHz
Voltage
0.2+0.2
0.1+0.1
0.2+0.2
0.3+0.4
Current via direct input
0.4+0.4
0.15+0.1
0.2+0.2
0.5+0.5
Active Power via direct input
0.5+0.5
0.2+0.1
0.3+0.2
0.6+0.5
Current via voltage input
0.2+0.2
0.1+0.1
0.2+0.2
0.3+0.4
Active power via voltage input
0.3+0.3
0.15+0.1
0.3+0.2
0.6+0.5
Just for AC Coupling
Frequency/Hz
45...65Hz
Voltage
0.05+0.05
Current via direct input
0.05+0.05
Active Power via direct input
0.07+0.04
Current via voltage input
0.05+0.05
Active power via voltage input
0.07+0.07
With DC coupling you have to use the first table!
Accuracies based on:
1. sinusoidal voltages and currents
2. ambient temperature (23
±
3)°C, no additional heating or cooling (i.e. by sunlight or current
of air)
3. warm up time 1h
Summary of Contents for LMG450
Page 1: ...4 Channel LMG450 Power Meter User manual Status 2009 02 17...
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 8: ......
Page 10: ......
Page 18: ......
Page 22: ......
Page 36: ......
Page 76: ......
Page 102: ......
Page 112: ......
Page 118: ......
Page 124: ......
Page 230: ......
Page 240: ......
Page 292: ......