Z E S ZIMMER Electronic Systems GmbH Chapter 10
201
10.2.10.14.2 :IUPDate
SACT
SCPI:
:SENSe:WAVeform:IUPDate/nquery/ [<NRi>]
SHORT: SACT/nquery/ [<NRi>]
ID:
n/a
Mode:
All
Type:
n/a
Suffix:
n/a
Value:
n/a
List:
n/a
Unit:
n/a
Group:
optional [<NRi>], 0=A, 1=B, ...
Requests new information about the scope data. Before this command you should set
‘:MEMory:FREeze ON’. After this command you can use ‘:SENS:WAV:SATR’,
‘:SENS:WAV:SBTR’, ‘:SENS:WAV:SSAM’ and ‘:SENS:WAV:SCTR’.
10.2.10.14.3 :SATRigger?
SATR?
SCPI:
:SENSe:WAVeform:SATRigger? /qonly/ [<NRi>]
SHORT: SATR? /qonly/ [<NRi>]
ID:
n/a
Mode:
All
Type:
long int
Suffix:
n/a
Value:
n/a
List:
n/a
Unit:
n/a
Group:
optional [<NRi>], 0=A, 1=B, ...
Reads how many sample values are available after the trigger event. See also
‘:SENS:WAV:IUPD’ for further information.
10.2.10.14.4 :SBTRigger?
SBTR?
SCPI:
:SENSe:WAVeform:SBTRigger? /qonly/ [<NRi>]
SHORT: SBTR? /qonly/ [<NRi>]
ID:
n/a
Mode:
All
Type:
long int
Suffix:
n/a
Value:
n/a
List:
n/a
Unit:
n/a
Group:
optional [<NRi>], 0=A, 1=B, ...
Reads how many sample values are available before the trigger event. See also
‘:SENS:WAV:IUPD’ for further information.
10.2.10.14.5 :SCTRigger?
SCTT?
SCPI:
:SENSe:WAVeform:SCTRigger? /qonly/
SHORT: SCTT? /qonly/
ID:
n/a
Mode:
All
Type:
long int
Suffix:
n/a
Value:
0...2
31
-1
List:
n/a
Unit:
n/a
Group:
n/a
Reads the number of the sample value at the trigger. See also ‘:SENS:WAV:IUPD’ for further
information. The sample values of the instrument are counted. At the end of each cycle this
Summary of Contents for LMG450
Page 1: ...4 Channel LMG450 Power Meter User manual Status 2009 02 17...
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 8: ......
Page 10: ......
Page 18: ......
Page 22: ......
Page 36: ......
Page 76: ......
Page 102: ......
Page 112: ......
Page 118: ......
Page 124: ......
Page 230: ......
Page 240: ......
Page 292: ......