Interfaces (IEEE option L45-O1) LMG450
182
10.2.9 :MEMory
commands
:CALCulate
:DISPlay
:FETCh
:FORMat
:INITiate
:INPut
:INSTrument
:MEMory
→
:FREeze
:READ
:SSIZe
:SENSe
:SOURce
:STATus
:SYSTem
:TRIGger
10.2.9.1 :FREeze
FRZ
SCPI:
:MEMory:FREeze <NRi>
SHORT: FRZ <NRi>
ID:
n/a
Mode:
All
Type:
long int
Suffix:
n/a
Value:
0, 1
List:
n/a
Unit:
n/a
Group:
n/a
Freezes the scope RAM. The scope has too much memory so it can’t be copied each cycle into
a separate buffer. For this reason you should set FRZ to ON when you want to readout the
sample values of the scope. Parameter:
‘ON’ or ‘1’ activates the freeze mode
‘OFF’ or ‘0’ deactivates the freeze mode [*RST default value]
10.2.9.2 :SSIZe
GMEM
SCPI:
:MEMory:SSIZe? /qonly/
SHORT: GMEM? /qonly/
ID:
n/a
Mode:
All
Type:
long int
Suffix:
n/a
Value:
0, 2
16
, 2
22
, 2
23
List:
n/a
Unit:
n/a
Group:
n/a
Reads the size of the sample value memory.
10.2.10 :SENSe commands
10.2.10.1 :AINPut
:CALCulate
:DISPlay
:FETCh
:FORMat
:INITiate