Z E S ZIMMER Electronic Systems GmbH Chapter 13
255
13.3 Measuring
channels
13.3.1 Sampling
The sampling is done at all channels with about 50kHz per channel.
13.3.2 Ranges
Voltage ranges
Rated range value / V
6
12.5
25
60
130
250
400
600
Measurable TRMS value / V
7.2
14.4
30
60
130
270
560
720
Permissible peak value / V
12.5
25
50
100
200
400
800
1600
Overload capability
600V continuously, 1500V for 1s
Input resistance
1M
Ω
, 10pF
Capacitance against earth
280pF
Common mode error (measured with 100V common mode signal at different frequencies in the
6V range without HF-Rejection filter):
50Hz
0V
5kHz
<0.01V
10kHz
<0.02V
20kHz
<0.05V
100kHz
<0.12V
Common mode error (measured with 100V common mode signal at different frequencies in the
6V range with HF-Rejection filter):
50Hz
0V
5kHz
<0.01V
10kHz
<0.02V
20kHz
<0.01V
100kHz
0V
Note!
The ‘Measurable TRMS value’ is the biggest TRMS value which can be measured. That
does not mean that it is allowed to measure with that value, if any security standards
define other values!
Current ranges
Rated range value / A
0.6
1.2
2.5
5
10
16
Measurable TRMS value / A
1.3
2.6
5.2
10
18
18
Permissible peak value / A
1.875
3.75
7.5
15
30
60
Overload capability
18A continuously, 50A for 1s, 150A for 20ms
Input resistance Ri
2m
Ω
Capacitance against earth
10pF
Summary of Contents for LMG450
Page 1: ...4 Channel LMG450 Power Meter User manual Status 2009 02 17...
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 8: ......
Page 10: ......
Page 18: ......
Page 22: ......
Page 36: ......
Page 76: ......
Page 102: ......
Page 112: ......
Page 118: ......
Page 124: ......
Page 230: ......
Page 240: ......
Page 292: ......