KAM

®

 LRW

 

LOW RANGE WATERCUT

User Manual

LRWMANUAL 0321

PTB 08 ATEX 1026

II 2 G Ex db IIB T6 Gb 

IECEx PTB 19.0048

Ex db IIB T6 Gb

TEL 

+1 713 784 0000  

FAX 

+1 713 784 0001  

Email

 [email protected] 

KAM CONTROLS, INC.

  

3939 Ann Arbor Drive 

Houston, Texas 77063 USA

www.KAM.com

An ISO 9001 certified company

Summary of Contents for LRW

Page 1: ...UAL 0321 PTB 08 ATEX 1026 II 2 G Ex db IIB T6 Gb IECEx PTB 19 0048 Ex db IIB T6 Gb TEL 1 713 784 0000 FAX 1 713 784 0001 Email Sales Kam com KAM CONTROLS INC 3939 Ann Arbor Drive Houston Texas 77063 U...

Page 2: ...2 Inline calibration and verification 26 Calibration with lab results 28 Process parameter values 29 Density input reconfiguration 30 Changing temperature units 32 Setting alarm setpoints 33 Setting a...

Page 3: ...Insertion LRW for pipelines 3 8 Insertable retractable LRW for pipelines 3 and up 2 3 or 4 flanged seal housing or 2 MNPT process connection Water 0 377 Temp 28 278 C www KAM com 1 713 784 0000 sales...

Page 4: ...compensate for any change in density across the entire API gravity range Compensation is based on a density input from a mass flow meter or densitometer and all compensation is automatic This is para...

Page 5: ...Accuracy 0 03 up to 3 water 0 05 from 3 5 water 0 1 from 5 10 water Repeatability 0 01 Resolution 0 01 Minimum water detection 100 ppm Outputs 4 20 mA Alarm relay RS232 RS485 Input 4 20 mA for densit...

Page 6: ...ult with the KAM factory INSTALLATION FLOW AND LOCATION REQUIREMENTS PLEASE NOTE The LRW should always be installed per KAM recommendations including the mixing requirement calculated per customer pro...

Page 7: ...units will be shipped partially assembled LRW unit can be wired prior to insertion into the spool for convenience but this is not required For 2 units FIG 3 1 1 Grease around the O ring below the smal...

Page 8: ...e user s responsibility to ensure that the LRW sensor be placed at the most representative point within the flow profile see flow recommendations on page 5 of this manual The LRW sensor should be inse...

Page 9: ...Housing models MID Minimum Insertion Distance D1 Distance from the top of the valve to the pipe Lg Gasket Width Typical 0 134 WT Pipe Wall Thickness Ls Pipe ID x 0 25 Measurement Area Length Ds Pipe I...

Page 10: ...cking Collar and place a mark with a permanent marker or tape on the Shaft Do not use anything sharp to mark the shaft as this will create grooves that will damage the O rings in the seal housing FIG...

Page 11: ...be conducted in accordance with all regional and class requirements Water 0 377 Temp 28 278 C www KAM com 1 713 784 0000 sales kam com Made in the USA LRW Low Range Watercut Water 0 377 Temp 28 278 C...

Page 12: ...will accumulate Lid should be tightly screwed shut all conduits should be sealed and secured in accordance with regional and classification requirements and unused 3 4 NPT openings should be sealed i...

Page 13: ...D 232 TXD GND GND 4 20 4 20 POWER POWER CHS CHS CHS GND DOUT GND DENSITY IN Kam LED1 TP1 TP3 GND LED2 KAM LRW www kam com askanengineer kam com Phone 1 713 784 0000 MADE IN USA CAUTION LRW provides th...

Page 14: ...20mA 4 20mA 24VDC 12 Watts Power Supply RECEIVER INPUT LRW LRW POWER POWER 4 20mA 4 20mA Power Supply 24VDC 12 Watts RECEIVER INPUT WRONG The 4 20 mA and power supply cannot be the same or connected...

Page 15: ...LOOP POWERED ISOLATOR RECOMMENDED FIG 3 13 Recommended 4 20 mA Loop Isolators 1 ASI X756526 Loop Powered Analog Signal Isolator DIN Rail Slim Line Single Channel 2 ASI 451129 4 20 mA Loop Powered Ana...

Page 16: ...RELAY WIRING FIG 3 15 DIGITAL RELAY CONFIGURATION 24 VDC TYPICAL WIRING WRONG LRW DOUT Low Side High Side Relay GND WRONG Relay GND DOUT Low Side LRW High Side AC Voltage Low Side Relay High Side LRW...

Page 17: ...ND LRW 4 20mA INPUT OUTPUT 2 4 20mA Signal Splitter POWER INPUT Density Wiring Density meter provides power for the 4 20mA Input Density Meter DENSITY IN GND LRW 4 20mA OUTPUT VDC Power Supply OUTPUT...

Page 18: ...agram 232 TXD Red 232 RXD White GND Black NOTE To avoid repeatedly opening electronics enclosure during diagnostics it is recommended operators permanently wire an RS232 connection to the instrument w...

Page 19: ...LRWMANUAL 0321 KAM CONTROLS INC 18 I N S TA L L AT I O N C O N T I N U E D WIRING CONTINUED FIG 3 20 RS485 WIRING DIAGRAM 485 485 RS485 Wiring Diagram TDA DATA TDB DATA RS485 Converter 2 Wire LRW...

Page 20: ...e boards unscrew the cover from the Explosion Proof electronics enclosure unscrew the LCD Display plate and carefully unplug the LCD connector 2 Connect the other end of RS232 cable to the serial port...

Page 21: ...h below Baud 115200 Parity None Data Bits 8 Stop Bits 1 Hardware Flow Control None Port Select port number assigned to your serial port or USB port connected to converter You can find this information...

Page 22: ...NC 21 K A M L R W O P E R AT I O N R E A LT E R M C O N T I N U E D 7 Click on the Send tab and select first four boxes under the EOL section FIG 4 4 8 The RealTerm software configuration is complete...

Page 23: ...o install Before proceeding install and configure RealTerm as per instructions on pages 19 21 of this manual Ensure your PC is connected to the LRW sensor via the supplied RS232 serial cable and USB t...

Page 24: ...for output data definitions If the output readings do not display in the RealTerm menu proceed to the RS232 Troubleshooting section on page 42 of this manual 2 Go to the Send tab type ostart on either...

Page 25: ...ncy kHz Column 4 minimum voltage Column 5 maximum voltage Column 6 minimum frequency kHz Column 7 frequency offset kHz Column 8 board temperature Range 0 to 70 C Column 9 probe temperature Range 0 to...

Page 26: ...L 0321 KAM CONTROLS INC 25 K A M L R W O P E R AT I O N R E A LT E R M C O N T I N U E D 4 Type ostop and click on Send ASCII to stop the continuous output Always do this before disconnecting FIG 4 8...

Page 27: ...water to be entered on 8 Realterm to calibrate the LRW See Calibration Example on page 27 9 Type lcal lower case L in either command port under the Send tab and click on Send ASCII The Calibration Mo...

Page 28: ...1 According to the averages the LRW is reading 0 11 high Current reading 3 0 To adjust the LRW the reading needs to be lowered by 0 11 3 0 0 11 2 89 This is the new value to be entered into the LRW K...

Page 29: ...according to API MPMS Chapter 8 1 waiting at least 15 minutes between sample draws 2 Each time a sample is drawn note or take a screenshot of the current LRW reading in the RealTerm window 3 Send the...

Page 30: ...ceeding install and configure RealTerm as per instructions on pages 19 21 of this manual Ensure your PC is connected to the LRW sensor via the supplied RS232 serial cable and USB to serial converter a...

Page 31: ...he LRW sensor via the supplied RS232 serial cable and USB to serial converter as per wiring diagram on page 12 To access the boards unscrew the cover from the Explosion Proof electronics enclosure uns...

Page 32: ...button associated with each port 1 2 Units for the density input range can either be API gravity or g cm3 Current input units are indicated under Process Parameter Values FIG 4 14 If the input units o...

Page 33: ...ial cable and USB to serial converter as per wiring diagram on page 12 To access the boards unscrew the cover from the Explosion Proof electronics enclosure unscrew the LCD Display plate and carefully...

Page 34: ...r command box and click on Send ASCII For example if the percent water where the alarm will activate is 3 you would enter alarm 3 The configured values will be displayed FIG 4 19 FIG 4 19 2 The hyster...

Page 35: ...r instructions on pages 19 21 of this manual Ensure your PC is connected to the LRW sensor via the supplied RS232 serial cable and USB to serial converter as per wiring diagram on page 12 To access th...

Page 36: ...the LCD Display plate and carefully unplug the LCD connector 1 Launch RealTerm Go to the Send tab type modbus on either command box and click on Send ASCII to check Modbus status The Modbus parameters...

Page 37: ...plied RS232 serial cable and USB to serial converter as per wiring diagram on page 12 To access the boards unscrew the cover from the Explosion Proof electronics enclosure unscrew the LCD Display plat...

Page 38: ...rm value Set high water percent alarm value alhy value Set alarm hysteresis lower limit for the alarm activation band alnc Set alarm mode to normally closed alno Set alarm mode to normally open date M...

Page 39: ...loth with oil solvent or part washer Preferred solvents include any petroleum solvent such as mineral spirits xylene toluene gasoline or diesel Do not use WD40 or other chemicals If you have a questio...

Page 40: ...he LCD connector from the terminal board FIG 6 2 and set LCD display plate aside 6 Remove the four 4 screws on the Terminal PCB with a Phillips screwdriver FIG 6 2 7 Carefully pull the Terminal Board...

Page 41: ...rily indicate proper voltage T R O U B L E S H O OT I N G C O N T I N U E D POWER CONNECTIONS CHECKING THE INPUT POWER 1 Use a voltmeter to check the POWER and POWER terminals and ensure that the volt...

Page 42: ...LRW supplies power for the loop Make sure the device that receives data from the LRW does not have a power source of its own switched on Also the 4 20 and 4 20 are isolated inputs from the chassis ea...

Page 43: ...1 through 4 on the previous section Loop Troubleshooting to access the terminal board of the LRW Check the RS232 wires connected to the LRW terminals for damage and ensure they are installed as per th...

Page 44: ...32 Bit Float 42008 Current Output Read 32 Bit Float 42009 Probe Temperature Read 32 Bit Float 42010 Adjusted Frequency Factory Debugging Read 32 Bit Float 42011 Offset Factory Debugging Read 32 Bit Fl...