Z E S ZIMMER Electronic Systems GmbH Chapter 5
101
•
Cycls allows you to define the minimum duration in times of the cycle time.
•
Because of this many influence factors the storing rate and the real duration (both
displayed) will vary if you change any of this parameters. In general you can say:
•
The bigger the memory (65536 words or 4194304 words) the bigger the record rate. If the
full record rate is reached, the duration will increase.
•
The fewer signals to be stored, the bigger the record rate. If the full record rate is reached,
the duration will increase.
•
The shorter the cycle time, the bigger the record rate. If the full record rate is reached, the
duration will increase.
If you transfer the sample values to a PC and you want to make evaluations in the frequency
domain (e.g. digital filtering or FFT), please make sure that the sampling (record) rate is bigger
than twice the bandwidth!
dot
The dot joiner connects two following pixels with a line. This function can be set to ‘on’
or ‘off’.
Please note that the grid is always displayed with integer numbers. The cursor positions are
calculated from the sample values and have not to fit to the grid.
5.4.7 Custom
menu
With
Custom you see the custom menu. Here you can setup your own menus, calculate
formulas and execute programs. See 4.4.3, ‘Custom menu’ and 4.4.4, ‘Script/Formula editor’.
5.5
Storage of values
First you have to change to the menu you want to print out or to store and press Print/Log
(exact handling see 11, ‘Logging of values to drives, printer and interfaces’). All the values you
see in this menu are printed out.
Summary of Contents for LMG450
Page 1: ...4 Channel LMG450 Power Meter User manual Status 2009 02 17...
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 8: ......
Page 10: ......
Page 18: ......
Page 22: ......
Page 36: ......
Page 76: ......
Page 102: ......
Page 112: ......
Page 118: ......
Page 124: ......
Page 230: ......
Page 240: ......
Page 292: ......