Z E S ZIMMER Electronic Systems GmbH Chapter 10
199
:TRIGger
:FILTer
:FINPut
:FLICker
:HARMonics
:INTegral
:SWEep
:TRANsient
:VOLTage
→
:RANGe
→
:AUTO
:WAVeform
:SCALe
:LINTern
:WIRing
[:UPPer]
:ZPReject
10.2.10.13.2.1 :AUTO
UAM
SCPI:
:SENSe:VOLTage:RANGe:AUTO <NRi>
SHORT: UAM <NRi>
ID:
n/a
Mode:
All
Type:
long int
Suffix:
1...4
Value:
0, 1
List:
n/a
Unit:
n/a
Group:
n/a
Reads and sets the status of the autorange function:
‘0’ or ‘MANUAL’ for manual range selection.
‘1’ or ‘AUTO’ for automatic range selection [*RST default value].
10.2.10.13.2.2 :LINTern?
UILS?
SCPI:
:SENSe:VOLTage:RANGe:LINTern?/qonly/
SHORT: UILS?/qonly/
ID:
n/a
Mode:
All
Type:
float
Suffix:
1...4
Value:
n/a
List:
n/a
Unit:
V
Group:
n/a
Reads a list with the available ranges. You get several float numbers which are separated by
colons ‘,’. The first number defines the number of following numbers.
10.2.10.13.2.3 [:UPPer]
URNG
RngU
SCPI:
:SENSe:VOLTage:RANGe[:UPPer] <NRf>
SHORT: URNG <NRf>
ID:
RngU
Mode:
All
Type:
float
Suffix:
1...4
Value:
n/a
List:
n/a
Unit:
V
Group:
n/a
Reads and sets the range for the voltage measurement. The parameter is the nominal value of
the range.
Summary of Contents for LMG450
Page 1: ...4 Channel LMG450 Power Meter User manual Status 2009 02 17...
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 8: ......
Page 10: ......
Page 18: ......
Page 22: ......
Page 36: ......
Page 76: ......
Page 102: ......
Page 112: ......
Page 118: ......
Page 124: ......
Page 230: ......
Page 240: ......
Page 292: ......