Transient mode (option L45-O5) LMG450
126
General
•
The settings of a group are valid for all channels of the group.
•
The channels of a group are OR combined. If at least one channel is triggered, the complete
group is triggered.
•
The groups are also OR combined. If one group is triggered, the other one is also triggered.
Figure 39 Measuring menu in transient mode
Signl
This defines which signal should be watched for the transient event. Following
settings are possible: i, u, p.
Sync+Filt Here you reach a sub-menu in which you can setup some values:
Sync
Selects the signal which is used for synchronisation. There are four
possible settings:
ChnU
The voltage signal is used
ChnI
The current signal is used
Line
The line signal is used
Extn
The signal at the external synchronisation jack is used.
Coupl
Selects how the voltage or current signal is coupled to the following
trigger stage. This setting has no influence to the measured signals!
Please refer to 5.1.2, ‘Group A/B tab’.
Filt
Here you can select if the filters in the signal path of voltage and current
are active or not. This filters are only in the signal way and don’t influence
the synchronisation. For the possible settings see 5.1.2, ‘Group A/B tab’
You have to leave this menu by pressing Enter or ESC. The settings for Coupl and
Sync you did in this menu are not used for the transient search itself but for
continuously measuring the TRMS and peak values of voltage and current. This
values are used to update the over and under range display in the status line.
Summary of Contents for LMG450
Page 1: ...4 Channel LMG450 Power Meter User manual Status 2009 02 17...
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 8: ......
Page 10: ......
Page 18: ......
Page 22: ......
Page 36: ......
Page 76: ......
Page 102: ......
Page 112: ......
Page 118: ......
Page 124: ......
Page 230: ......
Page 240: ......
Page 292: ......