Fundus Imaging
2660021169012 Rev. A 2017-12
CIRRUS HD-OCT User Manual
13-2
Fundus Imaging
Iris Imaging
Model 500
Model 5000
Methodology
Live OCT
Fundus technology
Line scanning
ophthalmoscope
Live fundus image
During alignment
During alignment and during
OCT scan
Optical source
Superluminescent diode (SLD),
840 nm
Superluminescent diode (SLD),
750 nm
Optical power
Nominal: 600+/–60
μ
W at the
cornea. Safety shutoff power:
maximum 825
μ
W at cornea
< 1.5 mW at the cornea
Field of view
36 degrees W x 22 degrees H
36 degrees W x 30 degrees H
Frame rate
> 1.7 Hz
> 20 Hz
Transverse resolution
45
μ
m (in tissue)
25
μ
m (in tissue)
Model 500 and Model 5000
Methodology
CCD camera
Resolution
1280 x 1024
Live iris image
During alignment
Summary of Contents for CIRRUS HD-OCT 500
Page 1: ...2660021156446 B2660021156446 B CIRRUS HD OCT User Manual Models 500 5000 ...
Page 32: ...User Documentation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 2 6 ...
Page 44: ...Software 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 3 12 ...
Page 58: ...User Login Logout 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 4 14 ...
Page 72: ...Patient Preparation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 5 14 ...
Page 110: ...Tracking and Repeat Scans 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 6 38 ...
Page 122: ...Criteria for Image Acceptance 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 7 12 ...
Page 222: ...Overview 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 9 28 ...
Page 256: ...Log Files 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 11 18 ...
Page 308: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 34 ...
Page 350: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 I 8 ...
Page 351: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 ...