CIRRUS HD-OCT User Manual
2660021169012 Rev. A 2017-12
Troubleshooting Power Problems
12-1
12 Routine Maintenance
NOTE: Except for the top fan filter, the CIRRUS HD-OCT has no user–replaceable parts. The
user must not attempt hardware repairs without consulting Zeiss service personnel. To do
so voids the instrument warranty. However, we may provide software updates that users
can install.
Troubleshooting Power Problems
This section assumes that the instrument will not power on. Troubleshooting your power
problem depends on whether or not you power the instrument through the optional power
table.
If Not Using the Optional Power Table
If you power the instrument directly from a wall outlet (not through the optional power
table), check the following to determine the source of the power problem:
1. Is there power available everywhere in your office?
• If not, there may be a localized power outage in your office or a general power
outage in your neighborhood.
• If so, proceed to Step 2.
2. Is the instrument power cord plugged in at both ends?
• If not, plug in the cord and try to power up the instrument.
If Using the Optional Power Table
If you are using the optional power table, the instrument is powered through it. Check the
following to determine the source of the power problem, in order:
1. Is there power available everywhere in your office?
• If not, there may be a localized power outage in your office or a general power
outage in your neighborhood.
• If so, proceed to Step 2.
2. Does the table have power (while the instrument does not)? You can test the table by
trying the lift.
• If the table has power, the power problem is within the instrument. First, check that
the instrument power cord is plugged in at the power table and at the instrument. If
the table does not have power, check that the table is plugged in at both the wall
outlet and at the table.
Summary of Contents for CIRRUS HD-OCT 500
Page 1: ...2660021156446 B2660021156446 B CIRRUS HD OCT User Manual Models 500 5000 ...
Page 32: ...User Documentation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 2 6 ...
Page 44: ...Software 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 3 12 ...
Page 58: ...User Login Logout 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 4 14 ...
Page 72: ...Patient Preparation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 5 14 ...
Page 110: ...Tracking and Repeat Scans 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 6 38 ...
Page 122: ...Criteria for Image Acceptance 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 7 12 ...
Page 222: ...Overview 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 9 28 ...
Page 256: ...Log Files 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 11 18 ...
Page 308: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 34 ...
Page 350: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 I 8 ...
Page 351: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 ...