Performance Verification Check
2660021169012 Rev. A 2017-12
CIRRUS HD-OCT User Manual
12-10
12.With Transparency set at 0% (opaque), use the triangles to drag the horizontal and
vertical scan line indicators until they intersect in the very center. This way, they should
cross in the center of the circle, which is the alignment target.
cáÖìêÉ=NOJS==_äìÉ=~åÇ=j~ÖÉåí~=iáåÉë=`ÉåíÉêÉÇ=çå=^äáÖåãÉåí=q~êÖÉí
13. Click Back at upper right (or double–click anywhere) to exit full screen mode.
14.Move the Transparency slider to 100% (transparent) and double–click again to make
the fundus image appear full screen. Now you are ready to evaluate the test.
Change Transparency
from 0% to 100%
Summary of Contents for CIRRUS HD-OCT 500
Page 1: ...2660021156446 B2660021156446 B CIRRUS HD OCT User Manual Models 500 5000 ...
Page 32: ...User Documentation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 2 6 ...
Page 44: ...Software 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 3 12 ...
Page 58: ...User Login Logout 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 4 14 ...
Page 72: ...Patient Preparation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 5 14 ...
Page 110: ...Tracking and Repeat Scans 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 6 38 ...
Page 122: ...Criteria for Image Acceptance 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 7 12 ...
Page 222: ...Overview 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 9 28 ...
Page 256: ...Log Files 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 11 18 ...
Page 308: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 34 ...
Page 350: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 I 8 ...
Page 351: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 ...