Appendix
2660021169012 Rev. A 2017-12
CIRRUS HD-OCT User Manual
C-10
The data was taken on a total of 26 eyes for the Wide Angle to Angle scan and 27 eyes for
the HD Angle scan measured by three operators on three CIRRUS 5000 devices and by a
single operator on the Visante OCT. The first qualified CIRRUS scan from any of the three
CIRRUS 5000 devices was used for comparison with the first qualified Visante scan.
Table 7 shows the results of the repeatability and reproducibility analyses for CIRRUS
CIRRUS 5000. Table 8 shows the mean difference between CIRRUS 5000 and Visante OCT.
Negative differences mean that the measurement on CIRRUS is smaller than the Visante
OCT.
Summary of Contents for CIRRUS HD-OCT 500
Page 1: ...2660021156446 B2660021156446 B CIRRUS HD OCT User Manual Models 500 5000 ...
Page 32: ...User Documentation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 2 6 ...
Page 44: ...Software 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 3 12 ...
Page 58: ...User Login Logout 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 4 14 ...
Page 72: ...Patient Preparation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 5 14 ...
Page 110: ...Tracking and Repeat Scans 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 6 38 ...
Page 122: ...Criteria for Image Acceptance 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 7 12 ...
Page 222: ...Overview 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 9 28 ...
Page 256: ...Log Files 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 11 18 ...
Page 308: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 34 ...
Page 350: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 I 8 ...
Page 351: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 ...