In summary, 282 subjects were qualified as normal subjects, enrolled in the NDB study,
and included for the RNFL database. Data from the same scans was analyzed and
evaluated to determine the normal range of ONH parameters. The normative database
does not have subjects with refractive errors outside the –12.00 D to +8.00 D range.
Therefore, the analysis below should be applied with caution to subjects with refractive
errors outside the –12.00 D to +8.00 D range.
Scan Selection Criteria
The scans were reviewed for image quality as part of the original study. One best quality
scan for each scan type was chosen for each subject per eye. Scans having the following
characteristics were excluded from the normative database:
• Signal Strength of 5 or lower.
• Large eye motion during image acquisition, resulting in a saccade that was within the
central 80% of the scan area.
• Area of data loss greater than 10% at the edge of the scan area.
• Presence of floater(s) obscuring macular area on Macular cube scans or measurement
circle on Optic Disc cube scans.
Scans were reviewed again after processing with the optic nerve head analysis algorithm,
to ensure that no floaters impacted the optic nerve head region, and to ensure that the
optic nerve head data was within the axial field of view of the scan. For three eyes a new
scan was selected to ensure acceptable ONH as well as RNFL results.
In practice, clinicians and operators should quantitatively and qualitatively review scans
before comparing them to normal ranges.
Data Analysis
Data was analyzed using the proprietary optic nerve head analysis algorithm to obtain five
main summary parameters: Disc Area (mm
2
), Average Cup to Disc Ratio (CDR), Vertical Cup
to Disc Ratio, Rim Area (mm
2
), and Cup Volume (mm
3
).
Summary of Contents for CIRRUS HD-OCT 500
Page 1: ...2660021156446 B2660021156446 B CIRRUS HD OCT User Manual Models 500 5000 ...
Page 32: ...User Documentation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 2 6 ...
Page 44: ...Software 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 3 12 ...
Page 58: ...User Login Logout 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 4 14 ...
Page 72: ...Patient Preparation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 5 14 ...
Page 110: ...Tracking and Repeat Scans 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 6 38 ...
Page 122: ...Criteria for Image Acceptance 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 7 12 ...
Page 222: ...Overview 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 9 28 ...
Page 256: ...Log Files 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 11 18 ...
Page 308: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 34 ...
Page 350: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 I 8 ...
Page 351: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 ...