Safety
2660021169012 Rev. A 2017-12
CIRRUS HD-OCT User Manual
1- 8
CAUTION:
Do not use the NAS device or the instrument with an extension cord or a
power strip (multiple portable socket outlet). For additional safety, do not plug the NAS
device and the instrument into the same wall outlet. Failure to observe this instruction
could result in electrical shock to the patient and/or examiner.
Printer Safety
WARNING: Except when powering the printer through an isolation transformer in the USB
configuration, peripheral devices such as printers must be placed at least 1.5 meters (4.9
feet) away from the patient, such that the patient cannot touch a peripheral device with
any part of his or her body while being examined. In addition, the instrument operator
must not attempt to touch the patient and a peripheral device at the same time while
examining the patient. Failure to observe this warning could result in electrical shock to the
patient and/or examiner. Use of a printer in a wireless configuration enables you to observe
this warning more easily.
WARNING: When using the printer in the USB configuration, you must power the printer
through an isolation transformer. Failure to observe this warning could result in electrical
shock to the patient and/or examiner. To do so, you must use a special power cable. In
North America, the required cable has an IEC–320–14 connector on one end and a NEMA
S–15R connector on the other end. This cable is included in the accessory kit shipped with
the instrument.
WARNING: To directly connect a printer to the CIRRUS HD-OCT using a network patch cord
(UTP cable), only use an unshielded RJ–45 connector. Use of a shielded network patch
cord will ground the printer through the CIRRUS HD-OCT, which could result in electrical
shock to the patient and/or examiner. It could also invalidate the system safety approval. In
this configuration, the printer must be placed at least 1.5 m away from the patient.
CAUTION:
If you use a non–approved device or if you connect it incorrectly—for
example, by plugging the printer into the wall while using a USB connection, or by using a
shielded network (UTP) cable—you could invalidate the system safety approval.
CAUTION:
We strongly recommend you use peripheral devices supplied or approved by
Zeiss, when available, because they will have been tested to work with the instrument. If
you do use a peripheral device that conforms with the requirements in this section but is
not supplied by Zeiss, do not install any unapproved third party software on the instrument.
Installation of any unapproved software, including drivers, could degrade the performance
of the instrument and/or lead to corrupted diagnostic or therapeutic information and may
void the instrument warranty.
CAUTION:
Do not use the printer or the instrument with an extension cord or a power
strip (multiple portable socket outlet). For additional safety, do not plug the printer and the
instrument into the same wall outlet. Failure to observe this instruction could result in
electrical shock to the patient and/or examiner.
Summary of Contents for CIRRUS HD-OCT 500
Page 1: ...2660021156446 B2660021156446 B CIRRUS HD OCT User Manual Models 500 5000 ...
Page 32: ...User Documentation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 2 6 ...
Page 44: ...Software 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 3 12 ...
Page 58: ...User Login Logout 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 4 14 ...
Page 72: ...Patient Preparation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 5 14 ...
Page 110: ...Tracking and Repeat Scans 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 6 38 ...
Page 122: ...Criteria for Image Acceptance 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 7 12 ...
Page 222: ...Overview 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 9 28 ...
Page 256: ...Log Files 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 11 18 ...
Page 308: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 34 ...
Page 350: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 I 8 ...
Page 351: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 ...