Appendix
2660021169012 Rev. A 2017-12
CIRRUS HD-OCT User Manual
A-22
were small. Based on R
2
values, age explains only 12% of the variability in some
parameters, while signal strength explains no more than 4%. Refractive error and axial
length explain less than 2% of variability. For these reasons, only age was used when
applying normative limits to Ganglion Cell Analysis parameters.
Age
The Ganglion Cell Analysis was developed utilizing 282 subjects aged 19 to 84. All
parameters decrease slowly with age.
shows the average Ganglion Cell
Analysis within the measurement annulus as a function of age.
cáÖìêÉ=^JNP==^îÉê~ÖÉ=d~åÖäáçå=`Éää=^å~äóëáë=^îÉê~ÖÉ=qÜáÅâåÉëë=~ë=~=cìåÅíáçå=çÑ=^ÖÉK
Ethnicity
The ethnicity breakdown of the Ganglion Cell Analysis is as follows:
43% Caucasians, 24% Asians, 18% African American, 12% Hispanic, 1% Indian, and 2%
mixed ethnicity. There are statistically significant differences among them in GCL + IPL
thickness as shown in the table below for Average GCL + IPL Thickness.
Table A-4 Ethnicity Breakdown of the Ganglion Cell Analysis
Results revealed that the mean difference in the average thickness between any two race
groups is within 4.3 mm. Subjects of European Descent have thinner ganglion cell plus
inner plexiform layer thickness on average, while subjects of Hispanic and Chinese descent
have thicker ganglion cell plus inner plexiform layer thickness (p < 0.001).
European Descent
Hispanic Descent
African Descent
Asian Descent
Average GCL + IPL Thickness (
μ
m)
84.1 (7.8)
88.8 (6.4)
86.3 (7.8)
89.4 (7.2)
Summary of Contents for CIRRUS HD-OCT 500
Page 1: ...2660021156446 B2660021156446 B CIRRUS HD OCT User Manual Models 500 5000 ...
Page 32: ...User Documentation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 2 6 ...
Page 44: ...Software 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 3 12 ...
Page 58: ...User Login Logout 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 4 14 ...
Page 72: ...Patient Preparation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 5 14 ...
Page 110: ...Tracking and Repeat Scans 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 6 38 ...
Page 122: ...Criteria for Image Acceptance 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 7 12 ...
Page 222: ...Overview 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 9 28 ...
Page 256: ...Log Files 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 11 18 ...
Page 308: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 34 ...
Page 350: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 I 8 ...
Page 351: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 ...