![background image](http://html.mh-extra.com/html/zeiss/cirrus-hd-oct-500/cirrus-hd-oct-500_user-manual_927494264.webp)
Performance Verification Check
2660021169012 Rev. A 2017-12
CIRRUS HD-OCT User Manual
12-8
5. Click Save and then either Finish or ID Patient to exit data acquisition. You will return
to the ID Patient screen.
6. Select the Performance Verification patient again and click Analyze.
7. In the Analysis screen, select the scan you just saved.
8. Select Macular Thickness Analysis in the right–hand column.
cáÖìêÉ=NOJQ==mÉêÑçêã~åÅÉ=sÉêáÑáÅ~íáçå=^å~äóëáë=pÅêÉÉå=fåáíá~ääó
9. When the scan loads, select OCT Fundus in the
Overlay
drop–down menu.
10.Set the Transparency slider to zero.
Select OCT Fundus overlay
Set Transparency to zero
Summary of Contents for CIRRUS HD-OCT 500
Page 1: ...2660021156446 B2660021156446 B CIRRUS HD OCT User Manual Models 500 5000 ...
Page 32: ...User Documentation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 2 6 ...
Page 44: ...Software 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 3 12 ...
Page 58: ...User Login Logout 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 4 14 ...
Page 72: ...Patient Preparation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 5 14 ...
Page 110: ...Tracking and Repeat Scans 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 6 38 ...
Page 122: ...Criteria for Image Acceptance 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 7 12 ...
Page 222: ...Overview 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 9 28 ...
Page 256: ...Log Files 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 11 18 ...
Page 308: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 34 ...
Page 350: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 I 8 ...
Page 351: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 ...