Performance Verification Check
12-11
Pass Condition
• Pass: After changing the Transparency to 100%, if both scan line indicators pass
partially or wholly through the center, the system passes the check. This means the
co–alignment of the fundus image and the OCT scan image is within the acceptable
range. Some examples of pass conditions appear below.
cáÖìêÉ=NOJT==bñ~ãéäÉë=çÑ=m~ëë=`çåÇáíáçåë
Failure Condition
• Fail: After changing the Transparency to 100%, if one or both scan line indicators pass
clearly within the black portion of the center, the system fails the check. Some
examples of failure conditions appear below.
Pass—both lines centered
Pass—horizontal line marginal, but OK
Pass—vertical line marginal, but OK
Summary of Contents for CIRRUS HD-OCT 500
Page 1: ...2660021156446 B2660021156446 B CIRRUS HD OCT User Manual Models 500 5000 ...
Page 32: ...User Documentation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 2 6 ...
Page 44: ...Software 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 3 12 ...
Page 58: ...User Login Logout 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 4 14 ...
Page 72: ...Patient Preparation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 5 14 ...
Page 110: ...Tracking and Repeat Scans 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 6 38 ...
Page 122: ...Criteria for Image Acceptance 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 7 12 ...
Page 222: ...Overview 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 9 28 ...
Page 256: ...Log Files 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 11 18 ...
Page 308: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 34 ...
Page 350: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 I 8 ...
Page 351: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 ...