CIRRUS HD-OCT User Manual
2660021169012 Rev. A 2017-12
Appendix
C-11
Table C-7: Repeatability and Reproducibility of CIRRUS HD-OCT 5000
Scan Type
Repeatability
Reproducibility
Parameter
Mean
SD
Limit
CV%
SD
Limit
CV%
Wide Angle to Angle
TISA 500 Nasal
0.151
0.025
0.071
16.801
0.030
0.083
19.614
TISA 750 Nasal
0.263
0.028
0.080
10.827
0.037
0.103
14.053
AOD 500 Nasal
0.439
0.075
0.209
17.012
0.081
0.226
18.365
AOD 750 Nasal
0.570
0.055
0.153
9.598
0.084
0.236
14.783
SSA Nasal
37.696
3.774
10.569
10.013
4.552
12.746
12.076
AC Angle Nasal
36.165
3.427
9.595
9.475
4.861
13.612
13.442
TISA 500 Temporal
0.150
0.027
0.076
18.169
0.032
0.090
21.368
TISA 750 Temporal
0.275
0.041
0.115
14.946
0.045
0.126
16.353
AOD 500 Temporal
0.445
0.080
0.223
17.893
0.090
0.252
20.243
AOD 750 Temporal
0.586
0.076
0.213
12.972
0.085
0.237
14.434
SSA Temporal
37.846
4.442
12.438
11.737
5.024
14.068
13.276
AC Angle Temporal
35.951
3.725
10.430
10.361
5.184
14.514
14.418
HD Angle
TISA 500 Nasal
0.158
0.017
0.048
10.764
0.022
0.061
13.786
TISA 750 Nasal
0.281
0.023
0.065
8.305
0.034
0.095
12.085
AOD 500 Nasal
0.461
0.053
0.148
11.496
0.066
0.185
14.332
AOD 750 Nasal
0.621
0.054
0.151
8.699
0.075
0.211
12.162
SSA Nasal
39.186
2.772
7.762
7.074
3.638
10.188
9.285
AC Angle Nasal
38.282
2.517
7.048
6.575
3.433
9.612
8.968
TISA 500 Temporal
0.161
0.020
0.057
12.685
0.026
0.072
16.033
TISA 750 Temporal
0.270
0.028
0.078
10.319
0.032
0.090
11.950
AOD 500 Temporal
0.475
0.064
0.179
13.415
0.075
0.209
15.699
AOD 750 Temporal
0.576
0.062
0.173
10.750
0.072
0.202
12.534
SSA Temporal
38.440
3.478
9.738
9.048
4.197
11.751
10.918
AC Angle Temporal
37.209
2.868
8.031
7.708
3.630
10.164
9.756
All statistics are estimated from two-way random-effect ANOVA model with random effects opera-
tor/device, eye and interaction between operator/device and eye.
Mean = Intercept of the ANOVA model
Repeatability Limit = 2.8 x Repeatability SD
Repeatability CV% = (Repeatability SD)/Intercept x 100%
Reproducibility SD = Square root of the sum of the operator/device variance, the interaction variance and
the residual variance
Reproducibility Limit = 2.8 x Reproducibility SD
Reproducibility CV% = (Reproducibility SD)/Intercept x 100%
Summary of Contents for CIRRUS HD-OCT 500
Page 1: ...2660021156446 B2660021156446 B CIRRUS HD OCT User Manual Models 500 5000 ...
Page 32: ...User Documentation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 2 6 ...
Page 44: ...Software 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 3 12 ...
Page 58: ...User Login Logout 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 4 14 ...
Page 72: ...Patient Preparation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 5 14 ...
Page 110: ...Tracking and Repeat Scans 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 6 38 ...
Page 122: ...Criteria for Image Acceptance 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 7 12 ...
Page 222: ...Overview 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 9 28 ...
Page 256: ...Log Files 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 11 18 ...
Page 308: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 34 ...
Page 350: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 I 8 ...
Page 351: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 ...