CIRRUS HD-OCT User Manual
2660021169012 Rev. A 2017-12
Analysis Related Report Options
10-5
cáÖìêÉ=NMJQ==jq^=mêáåí=léíáçåë=
The Macula Thickness Option, is the default option. You can reset any tab by selecting
Reset to Default in the lower right corner of the screen.
The Macula Multi–Slice Option
The Macula Multi–Slice Option, the second checkbox within the first tab of the Printout
Configuration screen will set up a report that, when saved and/or printed, yields a series of
pages, each of which show a set of selected Macula slices.
Which slices are shown, is determined by selecting the
second
tab of the Printout
Configuration window, the Macula Multi-slice Parameters tab,
.
When the Macula Multi–Slice option is selected, reports generated from macula cube
scans will include images of the central fast B-scan and adjacent B-scans. Such reports
include four fast B-scans per page. Selecting the number of B-scans to display and the
spacing between them, determines how many pages the report will be.
cáÖìêÉ=NMJR==j~Åìä~=jìäíáÓpäáÅÉ=m~ê~ãÉíÉêë
The Central Region is comprised of the central 1 mm (1000
μ
m) of the cube. This is the
equivalent of 500
μ
m above and 500
μ
m below the central B-scan. The Mid–Regions are
comprised of the next 1.0 mm above and below the Central Region. The Outer Regions are
the final 1.5 mm of area above and below the Mid Regions. These three regions add up to
the 6 mm height of the scan box and are equivalent to the ETDRS grid spacing in the
vertical direction.
Summary of Contents for CIRRUS HD-OCT 500
Page 1: ...2660021156446 B2660021156446 B CIRRUS HD OCT User Manual Models 500 5000 ...
Page 32: ...User Documentation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 2 6 ...
Page 44: ...Software 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 3 12 ...
Page 58: ...User Login Logout 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 4 14 ...
Page 72: ...Patient Preparation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 5 14 ...
Page 110: ...Tracking and Repeat Scans 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 6 38 ...
Page 122: ...Criteria for Image Acceptance 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 7 12 ...
Page 222: ...Overview 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 9 28 ...
Page 256: ...Log Files 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 11 18 ...
Page 308: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 34 ...
Page 350: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 I 8 ...
Page 351: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 ...