cáÖìêÉ=UJNQ=^Çî~åÅÉÇ=omb=^å~äóëáë=Ó=pÅêÉÉå=O
On Screen 2 (
), the sub-RPE illumination segmentation is overlaid on the RPE
Elevation Map. The sub-RPE illumination segmentation is shown with an outline. This
image shows the fovea location with a dot marking and a circle corresponding to 3 mm
and 5 mm diameter circles centered on the fovea. This image is superimposed on top of the
Fundus image.
Use the caliper to measure distances on the RPE Profile Map.
NOTE: Caliper measurements are not saved!
Calculated values are of the current scan minus the prior, as well as the percent change in
elevations and illumination since the previous scan.
The volume of RPE elevations, as well as the area and distance of the sub-RPE illumination
around and from the Fovea respectively, are shown. If no sub-RPE illumination is seen in
the circle, the value shown is 0.0. A warning message will appear if the 5 mm circle
extends outside of the scan window or if the scan is too low in the B-scan window.
Advanced RPE Analysis may occasionally identify RPE elevations or areas of sub-RPE
illumination in normal subjects. In a post-hoc analysis of 115 subjects from the diversified
normative database, the software identified RPE elevations in the 5 mm circle in 2.6% of
the subjects, with a mean of 0.006 mm
2
for area and 0.0002 mm
3
for volume. In the same
analysis, the software identified areas of sub-RPE illumination in 6.1% of the subjects, with
a mean area of 0.08 mm
2
.
NOTE: Calculated differences do not consider test-retest variability.
Summary of Contents for CIRRUS HD-OCT 500
Page 1: ...2660021156446 B2660021156446 B CIRRUS HD OCT User Manual Models 500 5000 ...
Page 32: ...User Documentation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 2 6 ...
Page 44: ...Software 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 3 12 ...
Page 58: ...User Login Logout 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 4 14 ...
Page 72: ...Patient Preparation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 5 14 ...
Page 110: ...Tracking and Repeat Scans 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 6 38 ...
Page 122: ...Criteria for Image Acceptance 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 7 12 ...
Page 222: ...Overview 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 9 28 ...
Page 256: ...Log Files 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 11 18 ...
Page 308: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 34 ...
Page 350: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 I 8 ...
Page 351: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 ...