8-17
IM 703155-01E
Checking the Output Waveform
8
Setting the Vertical (Y) and Horizontal (X) Axes
8.
Press the
Scale
soft key. The Scale menu appears.
Setting the Upper Limit of the Vertical Axis
9.
Press the
Y Top Value
soft key.
10.
Enter the upper limit of the vertical axis using
numeric keys & ENTER
.
Setting the Number of Orders of Magnitude of the Logarithmic Scale
9.
Press the
Y/Div
soft key.
10.
Enter the number of orders of magnitude of the logarithmic scale using
numeric
keys & ENTER
.
Executing the Auto Scale of the Vertical Axis
9.
Press the
Y Auto Scale
soft key. Auto scale is executed.
The vertical scale is set so that the accumulated value of the instantaneous power that
exceeds the average power, which is computed from the waveform data, fits on the
waveform display frame.
Setting the Display Begin Point of the Horizontal Axis (the Value at the Left Edge
of the Waveform Display Frame)
9.
Press the
X Begin Value
soft key.
10.
Enter the display begin point using
numeric keys & ENTER
.
Setting the Display End Point of the Horizontal Axis (the Value at the Right Edge of
the Waveform Display Frame)
9.
Press the
X End Value
soft key.
10.
Enter the display end point using
numeric keys & ENTER
.
Executing the Auto Scale of the Horizontal Axis
9.
Press the
X Auto Scale
soft key. Auto scale is executed.
The horizontal scale is set so that the amplitude of the instantaneous power that is
calculated from the waveform data fits on the waveform display frame.
8.3 CCDF Display