8-10
IM 703155-01E
Setting the Vertical (Y) and Horizontal (X) Axes
8.
Press the
Scale
soft key. The Scale menu appears.
9.
Press the
Y Unit
soft key. Possible choices are displayed.
10.
Use
rotary knob & SELECT
to select dBm or V.
Setting the Upper Limit of the Vertical Axis
11.
Press the
Y Top Value [V]
soft key.
If you selected dBm in step 10, press the
Y Top Value [dBm]
soft key.
12.
Enter the upper limit of the vertical axis using
numeric keys & ENTER
.
Setting the Voltage (or Power) Level per Grid
11.
Press the
Y Axis/Div [V]
soft key.
If you selected dBm in step 10, press the
Y Axis/Div[dBm]
soft key.
12.
Enter the voltage (or power) level per grid using
numeric keys & ENTER
.
Executing the Auto Scale of the Vertical Axis
11.
Press the
Y Axis Auto Scale
soft key. Auto scale is executed.
The vertical scale is set so that the maximum and minimum values of the waveform
data in the vertical direction fit on the waveform display frame.
8.2 Spectrum Display