14-5
IM 703155-01E
Troubleshooting, Maintenance, and Inspection
14
Code
Message
Description and Corrective Action
Reference
Section
1061
Not found user preset information.
Cannot execute preset when no preset information
12.2
is stored. Execute preset after saving the
preset information.
1062
Application is not registered.
A software application is not selected.
12.8
Run the software application after selecting it.
1100
Illegal character.
Use valid characters.
Chapters
10, 11
1101
Incorrect path name.
Specify a correct path or file name.
–
1102
File type is wrong.
Specify an extension that matches the file type.
7.3,
chapter 10
1103
Input path name is a directory. Please enter a file
–
Chapter 10
name.
1104
Input path name is a file. Please enter a directory
–
Chapter 10
name.
1105
No such file or directory.
–
Chapter 10
1106
The source and the destination are the same file.
–
10.7, 10.8
1107
File or directory already exists.
–
10.7, 10.8
1108
Cannot find directory.
–
Chapter 10
1109
Cannot open hidden file.
A period cannot be used as the first character of
Chapter 10
a file name.
1110
No such device.
A device that is not connected is specified.
Chapter 10
1111
Permission denied.
Read or write is not permitted.
–
1112
Device or resource busy.
–
–
1113
Text file busy.
–
–
1114
File too big.
–
–
1115
No space left on device.
–
–
1116
Read-only file system.
–
–
1117
File name too long.
Use a shorter file name.
Chapter 10
1118
Cannot load this file format.
–
–
1119
File open error.
Check the file.
Chapter 10
Clock Errors
Code
Message
Description and Corrective Action
Reference
Section
1305
Cannot find 10MHz input.
–
5.1, 13.3
1306
Cannot find 10MHz reference.
–
5.1, 13.3
1309
PLL does not lock.
If an external reference signal is selected, apply
a reference signal.
5.1, 13.3
14.2 Messages and Corrective Actions