13-7
IM 703155-01E
Rear Panel Auxiliary Input/Output
13
13.4 Event Output
CAUTION
Do not apply external voltage to the EVENT OUT terminal. If you do, the
VB8300 may malfunction.
The event data is 1-bit data synchronized to the 14-bit waveform data. When waveform
data is created, 2 event data bits synchronized to the I and Q waveform data are
created. An event output terminal is available for each output terminal set (OUT1 to 4).
Event Output Terminal (EVENT OUT)
Specifications
Item
Specifications
Connector type
BNC
Number of terminals
2 (for Suffix Code -0642 or -1282)
4 (for Suffix Code -0644 or -1284)
8 (for Suffix Code -0648 or -1288)
Output level
3.3 V CMOS
Output impedance
Approx. 50
Ω
Output Circuit
3.3 V CMOS
50
Ω
Timing Chart
Approx. 7.5 ns
Waveform output signal
(analog signal)
Event output signal
* When the low-pass filter is set to Through