14-10
IM 703155-01E
14.5 Replacing the Power Fuse
WARNING
• To prevent the possibility of fire, only use a fuse with the specified rating
(voltage, current, and type).
• Be sure to turn OFF the instrument and unplug the power cord before replacing
the fuse.
• Never short the fuse holder.
• The fuse is attached to only one side of the power line.
Fuse Rating
The VB8300 uses the following power fuse.
Maximum rated voltage
250 V
Maximum rated current
5 A
Type
Time lag
Standard
VDE/SEMKO/UL/CSA/SEV
Part Number
A1114EF
Replacement Procedure
Replace the power fuse according to the procedure below.
1.
Turn OFF the power switch.
2.
Remove the power cord from the power connector.
3.
Insert the tip of a flat-blade screw driver in the dented section of the fuse holder
on the power connector side, turn it in the direction of the arrow, and remove the
fuse holder.
4.
Remove the blown fuse that is attached to the tip of the fuse holder.
5.
Attach a new fuse to the fuse holder and attach the fuse holder back in the
original place.
Used fuse