8-29
IM AQ6370C-17EN
Program Function
1
2
3
4
5
6
7
8
App
Index
Program Command
Description
Instrument models on which the
command is valid, parameter
ranges, and supported variables.
ZOOM START WL
****.***nm
Sets the starting wavelength of
the display scale.
AQ6370C
50.000 to 1699.950 nm (0.001 step)
AQ6373/AQ6373B
1.000 to 1199.950 nm (0.001 step)
AQ6375/AQ6375B
600.000 to 2400.000 nm (0.001 step)
ZOOM START FREQ
***.****THz
Sets the starting frequency of
the display scale.
AQ6370C
11.5000 to 499.9950 THz
(0.0001 step)
AQ6373/AQ6373B
10.0000 to 856.4950 THz
(0.0001 step)
AQ6375/AQ6375B
60.0000 to 250.0000 THz
(0.0001 step)
ZOOM START WNUM
***.****cm
-1
Sets the starting wavenumber of
the display scale.
AQ6375/AQ6375B
2067.000 to 8333.000 cm
-1
(0.001 step)
ZOOM STOP WL
****.***nm
Sets the ending wavelength of
the display scale.
AQ6370C
600.050 to 2250.000 nm (0.001 step)
AQ6373/AQ6373B
350.050 to 1625.000 nm (0.001 step)
AQ6375/AQ6375B
1200.000 to 3000.000 nm
(0.001 step)
ZOOM STOP FREQ
***.****THz
Sets the ending frequency of the
display scale.
AQ6370C
176.5050 to 665.0000 (0.0001 step)
AQ6373/AQ6373B
250.0050 to 999.9000 (0.0001 step)
AQ6375/AQ6375B
125.0000 to 315.0000 (0.0001 step)
ZOOM STOP WNUM
***.****cm
-1
Sets the ending wavenumber of
the display scale.
AQ6375/AQ6375B
4167.000 to 10433.000 cm
-1
(0.001 step)
PEAK->ZOOM CENTER Sets the peak wavelength of the
waveform on the active trace.
Sets the wavelength to the display
scale’s center wavelength.
OVERVIEW DISPLAY OFF Sets OVERVIEW display during
ZOOM toOFF
OVERVIEW DISPLAY
LEFT
Sets OVERVIEW display during
ZOOM to the left side of the
waveform screen
OVERVIEW DISPLAY
RIGHT
Sets OVERVIEW display during
ZOOM to the right side of the
waveform screen
OVERVIEW SIZE LARGE Sets OVERVIEW display during
ZOOM to a large display
OVERVIEW SIZE SMALL Sets OVERVIEW display during
ZOOM to a small display
ZOOM INITIALIZE
Resets the display scale to the
initial state.
1: Generalized variables E, O, T, U, and V are valid only for the AQ6370C/AQ6370D/
AQ6375B.
8.3 Program Function Commands