8-13
IM AQ6370C-17EN
Program Function
1
2
3
4
5
6
7
8
App
Index
Error Encountered upon Execution of a Program
If an error occurs during the running of a program, an error number indicating the details
of the error is displayed in a window, and execution of the program is stopped.
Classification of Error Numbers
300 to 307 Errors caused by attempting a setting in manual operation which is disabled
320 to 326 Special command-related errors
340 to 347 Input/output-related errors
360 to 369 External memory-related errors
380, 381
Other errors
The above numbers can be read out using the SYSTem:ERRor[:NEXT]? command (see
section 7.6, “Instrument-Specific Commands”).
No. Message
Cause
Instrument models
on which the
command is valid
300 Parameter out of range
A variable value is out of range or is not
defined for a command that sets a parameter
using variables.
302 Scale unit mismatch
There is a difference between the Y-axis
scale of the active trace and the unit of
a parameter in the “LINE MKR 3 or 4”
command.
303 No data in active trace
Setting of the moving marker, a peak (or
bottom) search, or activation of the analysis
function was made with no data in the active
trace.
304 Marker value out of range Specified wavelength was out of the
sweep range in the moving marker or line
wavelength marker setting command.
305 No data in traces A or B No waveform data in traces A or B when
executing the “EDFA NF” command
306 Invalid data
Trace had no data when attempting to save it
to memory or to write it to FD/INT.
307 Unsuitable Write item
All data items were OFF at execution of
“WRITE DATA”.
320 Undefined variable
A command containing an undefined variable
was executed.
321 Variable unit mismatch
The unit of each variable does not agree
within a command containing two or more
variables.
322 Overflow
An overflow occurred in an arithmetic
operation.
323 Undefined marker
variable
A command containing a marker-value
variable was executed when no marker had
been displayed.
324 Invalid marker variable
A command containing the corresponding
variable was executed at a time other than
immedia tely after execution of a spectrum
width search, peak search, etc.
325 Undefined line number
GOTO command’s jumping destination is a
number other than 1 to 200.
326 F1 greater than F2
F1>F2 when the “IF F1 @@@@@ F2”
command was executed.
340 Printer paper empty
No printer paper.
AQ6370C/AQ6373/
AQ6375
341 Printer head up
No print is made because the printer’s head-
up lever is raised.
AQ6370C/AQ6373/
AQ6375
345 Option does not respond No response from an external device.
346 Option is not connected No external device is connected.
347 GP-IB2 not system
controller
System controller connected to the GP-IB2
port has been set to an external computer.
AQ6370C/AQ6373/
AQ6375
8.2 Executing a Program