7-40
IM AQ6370C-17EN
7.6 Instrument-Specific Commands
APPLication Sub System Commands
Overview
• This subsystem consists of data logging commands.
:APPLication:DLOGging:ETIMe?
Function
Queries the elapsed time of data logging (in
seconds).
Syntax
:APPLication:DLOGging:ETIMe?
Response <integer>
<integer> = Elapsed time [sec]
Example
:APPLICATION:DLOGGING:ETIME? ->
10220
Description • This is an overlap command.
• This command can be used on the AQ6370D/
AQ6370C/AQ6373B/AQ6375B.
• This command is invalid when data logging is
paused.
:APPLication:DLOGging:LPARameter:INT
erval
Function
Sets or queries the measurement interval of
data logging.
Syntax
:APPLication:DLOGging:LPARameter:IN
Terval<wsp><integer>[SEC]
:APPLication:DLOGging:LPARameter:IN
Terval?
<integer> = Measurement interval [sec] (0 =
SWEEP TIME)
Example
:APPLICATION:DLOGGING:LPARAMETER:IN
TERVAL 10
:APPLICATION:DLOGGING:LPARAMETER:IN
TERVAL? -> 10
Description • This command is invalid when data logging is
in progress.
• This is a sequential command.
• This command can be used on the AQ6370C/
AQ6370D/AQ6373B/AQ6375B.
ABORt Sub System Command
:ABORt
Function
Stops operations such as measurements and
calibration.
Syntax
ABORt
Example
ABORt
Explanation • Operations to be stopped are as follows:
:APPLication:DLOGging:STATe
:CALibration:ALIGn[:IMMediate]
:CALibration:ALIGn:EXTernal[:IMMe
diate]
:CALibration:ALIGn:INTernal[:IMMe
diate]
:CALibration:BANDwidth|BWIDth[:IM
Mediate]
:CALibration:WAVelength
:INITiate
:PROGram:EXECute
:HCOPy[:INITiate]
:HCOPy[:INITiate]:FUNCtion:CALCul
ate:LIST
:HCOPy[:INITiate]:FUNCtion:MARKer
:LIST
• This is an overlapping command.