8-21
IM AQ6370C-17EN
Program Function
1
2
3
4
5
6
7
8
App
Index
Program Command
Description
Instrument models on which the
command is valid, parameter
ranges, and supported variables.
START WL ****.***nm
Sets measurement-starting
wavelength.
AQ6370C/AQ6370D
50.000 to 1700.000 nm (0.001 step)
AQ6373/AQ6373B
1.000 to 1200.000 nm(0.001 step)
AQ6375/AQ6375B
600.000 to 2400.000 nm (0.001 step)
START FREQ ***.****
THz
Sets measurement-starting
frequency.
AQ6370C/AQ6370D
11.5000 to 500.0000 THz (0.0001 step)
AQ6373/AQ6373B
10.0000 to 856.5000 THz (0.0001 step)
AQ6375/AQ6375B
60.0000 to 250.0000 THz (0.0001 step)
START WNUM
****.***cm
-1
Sets measurement-starting
wavenumber.
AQ6375/AQ6375B
2067.000 to 8333.000cm
-1
(0.001 step)
STOP WL ****.***nm
Sets measurement-ending
wavelength.
AQ6370C/AQ6370D
600.000 to 2250.000 nm (0.001 step)
AQ6373/AQ6373B
350.000 to 1625.000 nm (0.001 step)
AQ6375/AQ6375B
1200.000 to 3000.000 nm (0.001 step)
STOP FREQ ***.****
THz
Sets measurement-ending
frequency.
AQ6370C/AQ6370D
176.5000 to 665.0000 THz
(0.0001 step)
AQ6373/AQ6373B
250.0000 to 999.9000 THz
(0.0001 step)
AQ6375/AQ6375B
125.0000 to 315.0000 THz
(0.0001 step)
STOP WNUM
****.***cm
-1
Sets measurement-ending
wavenumber.
AQ6375/AQ6375B
4167.000 to 10433.000cm
-1
(0.001 step)
Δλ->SPAN
Performs a spectrum width search
on the active trace, and sets the
results to the measuring span.
0nm SWEEP TIME **
sec
Sets sweep time used when
a sweep is made in a 0 nm
measuring span.
0 (MINIMUM), 1 to 50 (1step)
VIEW SCALE->
MEAS SCALE
Sets the current display conditions
to measuring conditions.
8.3 Program Function Commands